MACROTESTG3 - COMBIG3

EN - 84

Test on RCD without integral current breaking device (with accessory RCDX10)

Differential protection type (RCD):

AC ( ), A (

), B(

) – Generali (G), Selettivi (S) e Ritardati ( )

Voltage range Phase-Earth, Phase-Neutral:

100V

265V RCD type AC and A, 190V 265V RCD type B

Rated tripping currents (I

N): 0.3A

 10A

Frequency: 50/60Hz

 5%

RCD without integral current breaking device tripping current - (for General RCD only)

RCD type

I

N Range

I

N

[mA]

Resolution [mA]

Accuracy

AC, A, B

300mA

IN 10A (0.3

 1.1) I

N

 0.1I

N

- 0%, +5%I

N

Duration of RCD without integral current breaking device tripping time – TT/TN systems

x 1/2

x 1

x 2

x 5

AUTO

\ G

S G S

G S

G S

G S

G S

0.3A

1.0A

AC 999 999 999 999 999 999 200 250

50 150

 

310

A 999 999 999 999 999 999 200 250

50 150

 

310

B 999 999 999 999 999 999

310

1.1A

3.0A

AC 999 999 999 999 999 999 200 250

50 150

 

310

A 999 999 999 999 999 999 200 250

50 150

 

310

B 999 999 999 999 999 999

3.1A

6.5A

AC 999 999 999 999 999 999 200 250

50 150

 

310

A 999 999 999 999 999 999 200 250

50 150

 

310

B 999 999 999 999 999 999

6.6A

10.0A

AC 999 999 999 999 999 999 200 250

A 999 999 999 999 999 999

B

Table with duration of tripping time measurement [ms] - Resolution: 1ms, Accuracy:

(2.0

reading + 2digits)

Duration of RCD without integral current breaking device tripping time – IT systems

x 1/2

x 1

x 2

x 5

AUTO

\ G

S G S

G S

G S

G S

G S

0.3A

3.0A

AC 999 999 999 999 999 999 200 250

50 150

 

310

A

B

3.1A

6.5A

AC 999 999 999 999 999 999 200 250

50 150

 

310

A

B

6.6A

10.0A

AC 999 999 999 999 999 999 200 250

A

B

Table with duration of tripping time measurement [ms] - Resolution: 1ms, Accuracy:

(2.0

reading + 2digits)

Overall earth resistance without RCD tripping (Ra)

Voltage range Phase-Earth, Phase-Neutral:

100

265V, Frequency: 50/60Hz

5%

Global earth resistance in systems with Neutral

Range [

] Resolution

[

]

Accuracy

0.01

9.99

0.01

(5

reading + 0.1

)

10.0

199.9

0.1

(5

reading + 1

)

200

1999

1

(5

reading + 3

)

Ut LIM (UL): 25V or 50V, Maximum current: <15mA

Global earth resistance in systems without Neutral

Range [

] Resolution

[

]

Accuracy

1

1999

1

-0%, +(5.0

rdg + 3

)

Maximum current:

< ½ I

N

set ; Ut LIM (UL): 25V or 50V