HT instruments MACROTEST G3 User Manual
Page 12

MACROTESTG3 - COMBIG3
EN - 11
5.1.2. Automatic Power OFF for display and key sound
Touch the
icon. The screen to the side appears on the
display.
Move the slide bar reference of section "
" down/up to
turn off/on the Automatic Power OFF of the instrument after
a period of inactivity of 5 minutes.
Move the slide bar reference of section "
" down/up to
disable/enable the sound key when pressed. Confirm the
choices made and go back to the previous screen.
5.1.3. System
Touch the icon
to select the type of electrical system
(TT, TN or IT), of the mains frequency (50Hz, 60Hz), of the
limit value for contact voltage (25V, 50V) and rated voltage
value to be used for calculating the assumed short-circuit
current (see §). The screen to the side appears on the
display.
Move the slide bar references to select the options. Confirm
the choices made and go back to the previous screen.
5.1.4. Operator name entry
Touch the icon
to enter the name of the operator that
will be displayed in the header of each measurement
downloaded to PC. The screen to the side appears on the
display.
Set the desired name using the virtual keyboard (max 12
characters).
Confirm the settings or exit without saving.
5.1.5. System date/time setting
Touch the icon
to set the system date/time. The screen
to the side appears on the display.
Touch the "EU" icon for the European date/time system in the
format "DD/MM/YY hh:mm" or the "US" icon for the American
system in the format "MM/DD/YY hh:mm AM/PM".
Touch the up/down arrow keys to set the desired value.
Confirm the settings or exit without saving.
Current date/time is kept inside the instrument without
batteries for approximately 12 hours.