Precautions and safety measures – HT instruments MACROTEST G3 User Manual

Page 4

background image

MACROTESTG3 - COMBIG3

EN - 3

1. PRECAUTIONS AND SAFETY MEASURES

The instrument has been designed in compliance with directives IEC/EN61557 and
IEC/EN61010-1 relevant to electronic measuring instruments. Before and after carrying out
the measurements, carefully observe the following instructions:

 Do not carry out any voltage or current measurement in humid environments.
 Do not carry out any measurements in case gas, explosive materials or flammables are

present, or in dusty environments.

 Avoid any contact with the circuit being measured if no measurements are being

carried out.

 Avoid contact with exposed metal parts, with unused measuring probes, etc.
 Do not carry out any measurement in case you find anomalies in the instrument such

as deformations, breaks, substance leaks, absence of display on the screen, etc.

 Pay special attention when measuring voltages higher than 25V in special

environments (such as construction sites, swimming pools, etc.) and higher than 50V in
normal environments, since a risk of electrical shock exists.

 Only use original HT accessories.

The following symbols are used in this manual:

CAUTION: observe the instructions given in this manual; improper use could
damage the instrument, its components or create dangerous situations for the
operator.

High voltage danger: electrical shock hazard.

Double insulation

AC voltage or current

DC voltage or current

Connection to earth

The symbol indicates that the instrument must not be connected to systems with
phase-to-phase rated delta voltage higher than 415V.


1.1. PRELIMINARY

INSTRUCTIONS

 This instrument has been designed for use in the environmental conditions specified in

§ 10.4.1. Do not use in different environmental conditions.

 The instrument may be used for measuring and verifying the safety of electrical

systems. Do not use on systems exceeding the limit values specified in § 10.3.

 We recommend following the normal safety rules devised to protect the user against

dangerous currents and the instrument against incorrect use.

 Only the accessories supplied with the instrument guarantee compliance with safety

standards. They must be in good conditions and be replaced with identical models,
when necessary.

 Make sure the batteries are correctly installed.
 Before connecting the test leads to the circuit being measured, check that the desired

function has been selected.

This manual is related to the following products: