MACROTESTG3 - COMBIG3

EN - 15

10. At the end of the test, if the value of the measured

resistance is higher than the set limit, the screen to the
side is shown on the display.

The value is shown in red and the

symbol indicates

the negative result of the measurement. The "> 99.9

"

message indicates the instrument overload status.

Press the SAVE button or touch the

icon to save

the measurement (see § 7.1).


6.1.1. Anomalous situations
1. In AUTO or "

" modes if the instrument detects a

resistance lower than the set limit value but for which is it
not able to generate a current of 200mA, the screen to
the side is displayed.

The

symbol is shown on the display and the values

of the real test current are indicated in red.

2. If in

mode the instrument detects a resistance

higher than 2

 at its terminals, it resets the offset value

and displays a screen like the one to the side. The

icon is shown on the display to indicate the calibration
reset value (i.e. performing the operation with open
terminals).

3. In case the instrument detects that the calibrated

resistance is higher that the measured resistance, the
instrument gives out a long sound and displays a screen

like the one to the side. The

icon is shown on the

display to indicate the calibration reset value.