beautypg.com

HT instruments MACROTEST G3 User Manual

Page 45

background image

MACROTESTG3 - COMBIG3

EN - 44

6.4.6. Verify of protection against indirect contacts (TN system)
1. Select the options "TN", “25 or 50V”, “50Hz or 60Hz”

and the reference voltage in the general settings of the
instrument (see § 5.1.3).

Touch the

icon. The screen to the side appears on

the display.

Touch the lower icon. The following screen appears on
the display:

2. Move the left slide bar reference by selecting the

icon to execute the measurement.

Move the central slide bar reference by selecting the
"L-PE" option. The right slide bar reference is
automatically set in position

.


Confirm the choice by going back to the initial
measurement screen.

3. Move the slide bar reference by selecting the type of

protection (Fuse of type gG or aM or magnetothermal
MCB in curve B, C, K, D).

Touch the “In” field. The following screen appears on
the display

4.

Touch the

icon to zero the value in the In field and

use the virtual keyboard to set the value of the nominal
current of the protection within the values allowed by
the instrument (see § 6.4.4)

Confirm the choice by going back to the previous
screen.

Touch the icon in the lower right corner to set the
tripping time of the protection. The following screen
appears on the display

5. Move the slide bar reference by selecting the protection

tripping time between the options: 0.1s, 0.2s, 0.4s, 5s.

Confirm the choice by going back to the initial
measurement screen

This manual is related to the following products: