HT instruments MACROTEST G3 User Manual
Page 56

MACROTESTG3 - COMBIG3
EN - 55
8. At the end of the test, if the detected phase sequence is
correct, the instrument displays a screen like the one
shown to the side (result "1-2-3").
Press the SAVE button or touch the
icon to save
the measurement (see § 7.1).
9. At the end of the test, if the two detected voltages are
in phase (phase concordance between two distinct
three-phase systems), the instrument displays a
screen like the one to the side (result "1-1-").
10 At the end of the test, if the detected phase sequence is
not correct, the instrument displays a screen like the
one shown to the side (result "2-1-3").
Press the SAVE button or touch the
icon to save
the measurement (see § 7.1).