PST & PSS & PSH SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 1 ⎯

1. INTRODUCTION

In the modern automatic measurement system, communication

between equipments and computers is essential. The measured
procedures can be varied with users’ testing programs, therefore, the
programmable power supply can be operated remotely from an
instrument controller or computer across the RS232 interface (optional)
or GPIB (optional).

Interface selection and setup
The GPIB address can be changed in normal operation condition. Press
[SHIFT] key and [LOCAL] key on the front panel, in which the last
transmitting interface settings will be displayed. Select interface and
press [ENTER], then select the baud rate (or GPIB address) and press
[ENTER] to confirm the setting by using the knobs. Finally, select “save”
and press [ENTER] to store the setup.

2. CONNECTING THE PROGRAMMABLE POWER

SUPPLY VIA GPIB INTERFACE

The GPIB interface capabilities:
The GPIB interface of the programmable power supply corresponds to
the standard of IEEE488.1-1987, IEEE488.2-1992 and SCPI-1994. The
GPIB interface functions are listed as follows:
SH1(Source Handshake)

: The power supply can transmit multilane

messages across the GPIB.

AH1(Acceptor Handshake) : The power supply can receive multilane

messages across the GPIB.

T6(Talker)

: Talker interface function includes basic

talker, serial poll, and unaddress if MLA
capabilities, without talk only mode
function.

PST & PSS & PSH SERIES PROGRAMMABLE POWER SUPPLY

PROGRAMMER MANUAL

⎯ 2 ⎯

L4 (Listener)

: The power supply becomes a listener

when the controller sends its listen
address with the ATN (attention) line
asserted. The power supply does not
have listen only capability.

SR1 (Service Request)

: The power supply asserts the SRQ

(Service request) line to notify the
controller when it requires service.

RL1 (Remote/Local)

: The power supply responds to both the

GTL(Go to Local) and LLO(Local Lock
Out) interface messages.

PP0 (Parallel Poll)

: The power supply has no Parallel Poll

interface function.

DC1 (Device Clear)

: The power supply has Device clear

capability to return the device to power
on status.

DT0 (Device Trigger)

: The power supply has no Device Trigger

interface function.

C0 (Controller)

: The power supply can not control other

devices.

Notes for GPIB installation

When the programmable power supply is set up with a GPIB system,
please check the following things:

z

Only a maximum of 15 devices can be connected to a single GPIB
bus.

z

Do not use more than 20m of cable to connect devices to a bus.

z

Connect one device for every 2m of cable used.

z

Each device on the bus needs a unique device address. No two
devices can share the same device address.

This manual is related to the following products: