APPENDIX

53

Specification

Output Function

Sine, Square, Triangle

Amplitude Range 10Vp-p (into 50Ω load)
Impedance

50Ω ± 10%

Attenuator

−20dB ± 1dB x2

DC Offset

< −5V ~ >+5V (50Ω load)

Duty Range

20% ~ 80%, 2Hz~1MHz (Square Wave)

Duty Resolution

1% (Square Wave Only)

Main

Display

9 digits LED display
SFG-2004 0.1Hz

~

4MHz

SFG-2007

0.1Hz ~ 7MHz

SFG-2010 0.1Hz

~

10MHz

SFG-2020

1Hz ~ 20MHz

SFG-2104 0.1Hz

~

4MHz

SFG-2107

0.1Hz ~ 7MHz

SFG-2110 0.1Hz

~

10MHz

Sine/Square
Waveform Range

SFG-2120

1Hz ~ 20MHz

Triangle
Waveform Range

0.1Hz ~ 1MHz
(1Hz ~ 1MHz for SFG-2020/2120)

Resolution

0.1Hz (1Hz for SFG-2020/2120)

Stability ±20ppm
Accuracy

±20ppm

Frequency

Aging ±5ppm/year

Harmonic
Distortion

≥−55dBc, 0.1Hz ~ 200kHz
≥−40dBc, 0.2MHz ~ 4MHz
≥−30dBc, 4MHz ~ 20MHz
(Amplitude at maximum position
without attenuating 1/10 of any
combination setting, TTL/CMOS off)

Sine Wave

Flatness

< ± 0.3dB, 0.1Hz ~ 1MHz
< ± 0.5dB, 1MHz ~ 4MHz
< ± 2dB, 4MHz ~ 10MHz
< ± 3dB, 10MHz ~ 20MHz
(At the max amplitude relating to 1kHz)

Triangle Wave

Linearity

≥ 98%, 0.1Hz ~ 100kHz
≥ 95%, 100kHz ~ 1MHz

Symmetry

±1% of period + 4ns, 0.1Hz ~ 100kHz

Square Wave

Rise/Fall Time

≤ 25ns at maximum output (50Ω load)

Level

4 ± 1Vpp ~ 15 ± 1Vpp, adjustable

CMOS Output

Rise/Fall Time

≤ 120ns

Level

≥ 3Vpp

TTL Output

Fan Out

20 TTL Load

SFG-2000 Series User Manual

54

Rise/Fall Time

≤ 25ns

Sweep Rate

100:1 Ratio Max, adjustable (Note1)

Sweep Time

1 ~ 30s, adjustable (Note2)

Sweep
(SFG-2100
Series)

Sweep Mode

Linear, Logarithmic

Depth

0 ~ 100%

Frequency

400Hz (internal), DC ~ 1MHz (external)

Carrier BW

100Hz ~ 5MHz (−3dB)

Amplitude
Modulation
(SFG-2100
Series)

External
Modulating

Sensitivity: ≤ 10Vpp for 100%
modulation

Deviation

≥0 ~ ±50kHz, center at 1MHz

Frequency

400Hz fixed (internal), 1kHz fixed
(external)

Frequency
Modulation
(SFG-2100
Series)

External
Modulating

Sensitivity: ≤ 10Vpp for 10%
modulation (center at 1MHz)

Range

5Hz ~ 150MHz

Accuracy

Time Base Accuracy ± 1 count

Time Base

±20ppm (23±5°C) after 30min warmup

Max. Resolution

100nHz for 1Hz, 0.1Hz for 100MHz

Input Impedance 1MΩ/150pf

Frequency
Counter
(SFG-2100
Series)

Sensitivity

≤ 35mVrms, 5Hz ~ 100MHz
≤ 45mVrms, 100MHz ~ 150MHz

Store/Recall

10 Groups of Setting Memories

Power Source

AC115V/230V +10%, −15%, 50/60Hz

Operation
Environment

Indoor Use, Altitude Up to 2000m
Ambient Temperature 0 ~ 40°C
Relative Humidity ≤ 80%, 0 ~ 40°C
Relative Humidity ≤ 70%, 35 ~ 40°C
Install Category II / Pollution Degree 2

Storage
Environment

Temperature: −10 ~ 70°C
Humidity: ≤70%

Accessories

Instruction Manual x 1, Power Cord x 1
GTL-101 x 2 (SFG-2100 Series)
GTL-101 x 1 (SFG-2000 Series)

Dimension

107 (W) x 266 (H) x 293 (D)

General

Weight

Approx. 3.2kg (SFG-2100 Series)
Approx. 3.1kg (SFG-2000 Series)

Note1: In order to get the maximum sweep span, sweep time needs to be tuned.
Note2: If sweep time is too long, the stop frequency will stay at the maximum
frequency until the end of the sweep cycle.

This manual is related to the following products: