Tore/recall setting – GW Instek SFG-2000 Series User Manual

Page 21

background image

COUNTER

INPUT

41

The following table shows the relationship between

counter frequency, gate time, resolution, and display.

Input

Gate
time

Resolution Display

0.01s 100µHz 1.0000Hz
0.1s

10µHz

1.00000Hz

1s 1µHz

1.000000Hz

Gate time /
Resolution
table

1Hz

10s

100nHz

1.0000000Hz

0.01s

100µHz

10.0000Hz

0.1s

10µHz

10.00000Hz

1s

1µHz

10.000000Hz

10Hz

10s

100nHz

10.0000000Hz

0.01s

1mHz

100.000Hz

0.1s

100µHz

100.0000Hz

1s

10µHz

100.00000Hz

100Hz

10s

1µHz

100.000000Hz

0.01s

10mHz

1.00000kHz

0.1s

1mHz

1.000000kHz

1s

100µHz

1.0000000kHz

1kHz

10s

10µHz

1.00000000kHz

0.01s

10Hz

1.00000MHz

0.1s

1Hz

1.000000MHz

1s

100mHz

1.0000000MHz

1MHz

10s

10mHz

1.00000000MHz

0.01s

100Hz

10.0000MHz

0.1s

10Hz

10.00000MHz

1s

1Hz

10.000000MHz

10MHz

10s

100mHz

10.0000000MHz

0.01s

100Hz

100.0000MHz

0.1s

10Hz

100.00000MHz

1s

1Hz

100.000000MHz

100MHz

10s

100mHz

00.0000000MHz*

Over sign

* The OVER sign shows that the 100MHz
digit contains a real number but hidden

from the display. It turns On under the
following condition.

Input frequency: ≥100MHz

Gate Time: 10s

To view the 100MHz digit, switch to smaller
gate time such as 1s or 0.1s.

SFG-2000 Series User Manual

42

S

TORE/RECALL SETTING

STORE

6

Press the Shift key, then 6 to store

the current panel setting to internal
memories.

Memory range: 0 ~ 9 (10 set)

Store the
panel setting

The “Store 0” sign appears.

Enter the memory number, 0 ~ 9. (for example, 1)

The “done” sign appears.

RECALL

3

Press the Shift key, then 3 to recall a
panel setting stored in internal

memories.

Memory range: 0 ~ 9 (10 set)

Recall the
panel setting

The “recall 0” sign appears.

Enter the memory number, 0 ~ 9. (for example, 1)

The “done” sign appears and the panel setting changes
accordingly.