Tore/recall setting – GW Instek SFG-2000 Series User Manual
Page 21
COUNTER
INPUT
41
The following table shows the relationship between
counter frequency, gate time, resolution, and display.
Input
Gate
time
Resolution Display
0.01s 100µHz 1.0000Hz
0.1s
10µHz
1.00000Hz
1s 1µHz
1.000000Hz
Gate time /
Resolution
table
1Hz
10s
100nHz
1.0000000Hz
0.01s
100µHz
10.0000Hz
0.1s
10µHz
10.00000Hz
1s
1µHz
10.000000Hz
10Hz
10s
100nHz
10.0000000Hz
0.01s
1mHz
100.000Hz
0.1s
100µHz
100.0000Hz
1s
10µHz
100.00000Hz
100Hz
10s
1µHz
100.000000Hz
0.01s
10mHz
1.00000kHz
0.1s
1mHz
1.000000kHz
1s
100µHz
1.0000000kHz
1kHz
10s
10µHz
1.00000000kHz
0.01s
10Hz
1.00000MHz
0.1s
1Hz
1.000000MHz
1s
100mHz
1.0000000MHz
1MHz
10s
10mHz
1.00000000MHz
0.01s
100Hz
10.0000MHz
0.1s
10Hz
10.00000MHz
1s
1Hz
10.000000MHz
10MHz
10s
100mHz
10.0000000MHz
0.01s
100Hz
100.0000MHz
0.1s
10Hz
100.00000MHz
1s
1Hz
100.000000MHz
100MHz
10s
100mHz
00.0000000MHz*
Over sign
* The OVER sign shows that the 100MHz
digit contains a real number but hidden
from the display. It turns On under the
following condition.
•
Input frequency: ≥100MHz
•
Gate Time: 10s
To view the 100MHz digit, switch to smaller
gate time such as 1s or 0.1s.
SFG-2000 Series User Manual
42
S
TORE/RECALL SETTING
STORE
6
Press the Shift key, then 6 to store
the current panel setting to internal
memories.
Memory range: 0 ~ 9 (10 set)
Store the
panel setting
The “Store 0” sign appears.
Enter the memory number, 0 ~ 9. (for example, 1)
The “done” sign appears.
RECALL
3
Press the Shift key, then 3 to recall a
panel setting stored in internal
memories.
Memory range: 0 ~ 9 (10 set)
Recall the
panel setting
The “recall 0” sign appears.
Enter the memory number, 0 ~ 9. (for example, 1)
The “done” sign appears and the panel setting changes
accordingly.