Using Power Quality Measurements

Background

For typical power measurements, one channel is
used to measure voltage using a differential probe
and the other channel is used to measure current
using a current probe.

In the example below, the power quality of an AC
power source is tested.

WARNING

Ensure safe working practices are adhered to when
working with live voltages. Failure to do so could
lead to electric shock or loss of life.

Connection

Differential probe: Line and Neutral.
Current probe: Line.

Setup

1. Deskew the current and voltage

probes.

Page 7

2. With the power disconnected from the AC

power source, connect the differential voltage
probe to the Line and Neutral wires and the
current probe to the Line wire.

3. Connect the differential probe and current

probe to an input channel.

4. Configure the channel with the differential

probe to the following settings:

Probe

Voltage