GOS-6112 OSCILLOSCOPE

USER MANUAL

 45 

Figure 5-11(a)

Figure 5-11(b)

Triggering of Video signal

In the work concerned with TV, complex signals and containing video

signal, blanking pedestal signal, and synchronizing signal are often

measured.

Press the TRIG MODE pushbutton to set the TV position. The built-in

active TV-Sync-separator provides the separation of frame or line sync

pulses from the video signal. To trigger the oscilloscope at the vertical

(frame) rate, press the TV-V/TV-H pushbutton to set TV-V coupling. To

trigger the oscilloscope at the horizontal (line), press the TV-V/TV-H

pushbutton to set TV-H coupling. The figure 5-12(a) shows vertical

signal of TV-V and Figure 5-12(b) shows horizontal signal of TV-H.

GOS-6112 OSCILLOSCOPE

USER MANUAL

 46 

Figure 5-12(a) TV-V

Figure 5-12(b) TV-H

The polarity of the synchronization pulse is critical for the slope selection.

The figure 5-13(a) and 5-13(b) shows the examples of TV polarity

synchronization signals.

Figure 5-13(a) (–) Sync signal.

Figure 5-13(b) (+) Sync signal