beautypg.com

GW Instek GOS-6100 Series User Manual

Page 24

background image

GOS-6112 OSCILLOSCOPE

USER MANUAL

 41 

Comparing Frequency and phase (Single X-Y Operation)

To compare the frequency and phase between two signals by using the

X-Y mode. The X-Y waveform displays different amplitude, frequency,

and phase. The figure 5-7 shows a typical waveform made up of two

signals that are of the same frequency and amplitude, but approximate

45

o

out of phase.

To use the oscilloscope in the signal of X-Y mode, proceed the following

steps:

1. Connect the horizontal or X-axis signal to the CH1 input.

2. Connect the vertical or Y-axis signal to the CH2 input.

3. Set the CH1 off, and set the CH2 on.

4. Set the X-Y mode on by pressing and holding the button.

5. Set the TRIG SORCE button to CH1.

Use the HORIZONTAL POSITION control to adjust the X-axis.

Note: When high frequency signals are displayed in the X-Y

operation, note the frequency bandwidths and phase difference

between X and Y axis. Refer to “2. SPECIFICATION” section for

details.

Figure 5-7 Typical single X-Y display.

GOS-6112 OSCILLOSCOPE

USER MANUAL

 42 

Setting up Dual X-Y Operation

To use the oscilloscope in the dual X-Y mode, proceed the following steps:

1. Connect the horizontal or X-axis signal to the EXT (X) input.

2. Connect one of the vertical or Y-axis signal to the CH1 (Y1) input.

3. Connect either of the vertical or Y-axis signal to the CH2 (Y2) input.

4. Set the CH1 and CH2 on.

5. Set the ALT/CHOP/ADD button to CHOP mode.

6. Set the X-Y mode on by pressing and holding the button.

The figure 5-8 shows two X-Y waveforms in the dual X-Y mode.

Figure 5-8 Typical dual X-Y display

Setting Up delayed-sweep Operation

A delayed sweep is used to magnify any portion of a complex waveform

in the horizontal direction.

To display the delayed sweep operation, proceed the following steps:

1. Briefly pressing the MAIN/ALT/DELAY pushbutton to set to MAIN

time base of the horizontal mode. Effect triggering by main sweep and

set MAIN TIME/DIV control as desired.