5-3.Adjustments and checks

Trace Rotation Adjustment

Normally, when the trace is in parallel with the center horizontal graticule

line, there will be no need to adjust the TRACE ROTATION. If necessary,

adjust the TRACE ROTATION to make the baseline trace parallel to the

center horizontal graticule line by using a small straight-blade

screwdriver or alignment tool.

Probe Compensation

To minimize the distortion of measured waveforms, check the

compensation of your probes before using them. The probe compensation

should be checked periodically whenever the probes are moved to

different input channels.

1. Install the probes onto the oscilloscope (Press the BNC connector onto

the channel input and rotate the connector to lock it into place).

2. Set the probe slide switches to the ×10 position.

3. Briefly pressing the CH1/CH2 button to set the oscilloscope to channel

1 and channel 2.

4. Pressing and holding the P×10 button to set the indicated deflection

coefficient of the channel displayed in the readout as a symbol “P10”.

5. Attach the probe tips to the CAL connection in the front of the

oscilloscope.

6. Set the oscilloscope controls to display both channels:

VERTICAL: VOLTS/DIV 1V
COUPLING

DC

ALT/CHOP/ADD

CHOP

HORIZONTAL: MODE

MTB

TIME/DIV

0.5ms

TRIGGER:

MODE

ATO

SOURCE

CH1 or CH2

COUPLING

AC

SLOPE

GOS-6112 OSCILLOSCOPE

USER MANUAL

 36 

7. Observe the displayed waveform and compare them with the

waveforms shown in figure 5-2. If either probe needs to be adjusted,

proceed the step 8. If either probe does not need to be adjusted, proceed

the “Function Check”.

Figure 5-2 Typical Compensation Waveform

8.Adjust the probe by using a small insulated screwdriver. Slowly rotate

the adjustment control until the probe is properly compensated.