beautypg.com

GW Instek GOS-6100 Series User Manual

Page 25

background image

GOS-6112 OSCILLOSCOPE

USER MANUAL

 43 

2. Set the MAIN/ALT/DELAY button to ALT mode, and set the time

range of the DELAY TIME/DIVE control to be magnified.

The figure 5-9 below shows the main and delayed sweeps appear

simultaneously on the screen, and the window sector will appear.

3.Adjust the DELAY TIME control to move continuously the window

sector. Bring the window sector to the position to be magnified. Then,

the waveform between the window sector is magnified to occupy the

full area of the screen.

4.If necessary, press the TRACE SEP pushbutton to set the LED on. The

delayed sweep trace can be shifted vertically about ±3 divisions with

respect to the main sweep trace for the convenience of observation by

the POSITION control.

Figure 5-9 ALT Mode

Figure 5-10 Delay Mode

5.Set the DELAY mode to measure the magnified waveform only by

pressing the MAIN/ALT/DELAY button. The magnified waveform

shown in figure 5-10.

GOS-6112 OSCILLOSCOPE

USER MANUAL

 44 

Magnifying Waveform Events

Use the ×10 MAG pushbutton to view small portions of a waveform as

which is too far back from the starting point to view by using the

TIME/DIV control. To use the ×10 MAG button, proceed the following

steps:

1. Adjust the TIME/DIV to the fastest sweep that displays the event.

2. Rotate the HORIZONTAL POSITION control to move the event to

display on the center of screen.

3. Press the ×10 MAG button to switch the MAG LED on.

When above procedures have been done, the displayed waveform will be

expanded 10 times to the right and left from the center of screen as center

of expansion.

Operating HO (Hold off time) Control

When the measured signal is a complex waveform with two or more

repetition frequencies (period), triggering with the LEVEL control alone

may not be sufficient to attain a stable waveform display. In such a case,

the sweep can be stable synchronized to the measured signal waveform

by adjusting the HO (Hold off) time of the sweep waveform.

Figure 5-11(a) shows several different waveforms which overlapped on

the screen, marking the signal observation unsuccessful when the hold

off is set to minimum (the HO-LED is dark).

Figure 5-11(b) shows the undesirable portion of the signal is held off.

The same waveforms are displayed on the screen without overlapping.