beautypg.com

Technical specifications – GW Instek GOS-6103C User Manual

Page 4

background image

GOS-6103C OSCILLOSCOPE

USER MANUAL

3

7) Trigger signal output

The signal selected by the TRIGGER SOURCE is available. This

output may be used to connect to a frequency counter or other

instrument.

8) Panel setups lock

To avoid unintentional touch of the setting, the feature is extremely

useful for long term and repetitive measurements that used to be

performed under the same test condition of the oscilloscope setting.

9) LED indicator and buzzer alarm

The LED’s located in the front panel assist operation and indicated

additional information. Incorrect operation and the electrical end

position of control knobs are indicated by a warning beep.

10)SMD manufacturing technology

The instrument is built by using the most advanced SMD technology

so as to reduce the number of internal wiring and shorten the foil route

on the pc board. This will also greatly increase the high frequency

performance and the reliability of the product.

11)Auto Measurement

A built-in 6 digits universal counter is accurate within the range of

±0.01% and can measure frequency between 50Hz and 100MHz.

GOS-6103C OSCILLOSCOPE

USER MANUAL

4

2.TECHNICAL SPECIFICATIONS

Sensitivity 2mV~5V/DIV,11

steps

in 1-2-5 sequence

Sensitivity Accuracy

±3% (5 DIV at the center display )

Vernier Vertical Sensitivity

Continuously variable to 1/2.5 or less than
panel-indicated value

Frequency Bandwidth(-3dB) DC ~ 100MHz (2mV/DIV:DC ~ 20MHz)
Rise Time

3.5ns (2mV/DIV:17.5ns)

Signal Delay

Leading edge can be monitored

Maximum Input Voltage

400V (DC+AC) at 1kHz or less

Input Coupling

AC, DC, GND

Input Impedance

1MΩ±2% // 25pF approx.

Vertical Modes

CH1,CH2,DUAL(CHOP/ALT),ADD(DIFF mode can
be established when the CH2 is in the INV mode)

CHOP Repetition Frequency Approx. 250kHz.
Polarity (INV)

CH2 only

Bandwidth Limited

20MHz

VERTICAL

DEFLECTION

SYSTEM

Common-mode
Rejection Ration

50:1 or better at 50kHz

Trigger Modes

AUTO, NORM, TV

Trigger Source

CH1, CH2, LINE, EXT

Trigger Coupling

AC, DC, HFR, LFR

Trigger Slope

+/- polarity or TV sync polarity

Mode

Frequency INT

EXT

10Hz~20MHz 0.35DIV

50mVpp

AUTO

20MHz~100MHz 1.5DIV

150mVpp

DC~20MHz 0.35DIV

50mVpp

NORM

20MHz~100MHz 1.5DIV

150mVpp

Trigger Sensitivity

TV Sync

signal

1DIV

200mVpp

Trigger Level Range

INT : ± 4 DIV or more

EXT : ± 0.4 V or more

TV Sync

TV-V,TV-H

Max. External Input Voltage 400V (DC + AC peak) at 1kHz

TRIGGER

SYSTEM

External Input Impedance

1MΩ± 5% /

25pF approx.