beautypg.com

Troubleshooting guide, Figure 29 sensor cell and itm mating – Detcon PI-700 0-500ppb User Manual

Page 40

background image

Model PI-700 0-500ppb

PI-700 0-500ppb Instruction Manual

Rev. 0.0

Page 36 of 53

6. Troubleshooting Guide


Refer to the list of Failsafe Diagnostic features listed in Section 3.6.2 for additional reference in
troubleshooting activities. Listed below are some typical trouble conditions and their probable cause and
resolution path.

Plug-in Sensor

(Bottom View)

ITM

Bottom View

Locking Setscrew

Display Window

Figure 30 Sensor Cell and ITM Mating

Under-Range Fault

Probable Cause: Sensor Baseline drifted lower, Ambient Interference gasses reduced.
Repeat AutoZero. Use Zero Air or N

2

source.

Execute successful AutoSpan and verify adequate Sensor Life.
Check Raw counts in View Sensor Status. Should be close to 1,500 for high range and 3,000 for low range .
Replace plug-in sensor if fault continues.

Missing Sensor Fault

Probable Cause: Sensor is Missing, Failed Plug-in Sensor Electronics, or ITM I.S. Barrier Failure.
Make sure plug-in sensor is plugged in properly with correct orientation.
Swap plug-in sensor into another ITM to determine if plug-in sensor problem or ITM problem.
Replace the plug-in sensor if proven faulty.
Replace the ITM if proven faulty.

AutoSpan Calibration Faults – (Range, Stability and Clearing)

To clear any AutoSpan Calibration fault, the AutoSpan process must be completed successfully (Section3.4).
Use zero air gas after AutoSpan Calibration to avoid clearing fault from high background VOC levels.

Range Fault

Probable Causes: Failed Sensor, Cal Gas not applied or not applied at appropriate time, or problems w/ cal gas
and delivery
Check validity of span gas (check MFG date on cal gas cylinder).
Use proper cal gas regulators and tubing for highly absorbing or corrosive gasses.
Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or corroded).
Verify adequate Sensor Life.
Clean or replace the PID lamp.
Replace the plug-in VOC sensor.

Stability Fault

Probable Causes: Failed Sensor, empty or close to empty Cal Gas Cylinder, or problems with cal gas and
delivery