Figure 18 Removal of Cell Assembly


6. With fine tipped tweezers, grasp the lamp by placing the tips in the housing notch and gently pulling it out.

Be careful not to scratch the lamp lens or chip the edges.

Figure 19 Removal of Lamp


Cleaning the Lamp

Figure 20 Lamp cleaning


Wearing gloves, grab the lamp by the cylindrical glass body and clean the window by rubbing it against the
Polishing Pad. Use a circular motion and try to keep the window surface flat relative to the pad. Five seconds