3 replacement of itm, Replacement of itm, Figure 28 sensor cell and itm mating – Detcon PI-700 0-500ppb User Manual
Page 38
Model PI-700 0-500ppb
PI-700 0-500ppb Instruction Manual
Rev. 0.0
Page 34 of 53
d) Gently pull the plug-in sensor out of the ITM. Transfer to new plug-in sensor. Orient the new plug-in
sensor so that it matches with the female connector pins. When properly aligned, press the sensor in firmly to
make the proper connection.
Plug-in Sensor
(Bottom View)
ITM
Bottom View
Locking Setscrew
Display Window
Figure 29 Sensor Cell and ITM Mating
e) Thread the Splash Guard Adapter and Sample Chamber onto the ITM to a snug fit and tighten the locking
setscrew using the 1/16” Allen wrench.
f) Verify the gas type and range of the new sensor by checking in View Program Status. It is recommended
“AutoZero and AutoSpan functions be performed, as per Section 3.4 Calibration Mode, to match the new
intelligent plug-in sensor with the ITM.
5.3 Replacement of ITM
a) Remove the power source from the sensor assembly. Disconnect all sensor wire connections at the J-Box
taking note of the wire connections.
NOTE: It is necessary to remove power to the J-Box while changing the ITM in order to
maintain area classification.
b) Use a wrench and the wrench flats provided at the top section of the ITM and unthread the ITM until it can
be removed.
c) Use a 1/16” Allen wrench to release the locking setscrew that locks the ITM and Splash Guard Adapter
together (One turn will suffice - Do not remove setscrew completely).
d) Unthread and remove the Splash Guard Adapter and Sample Chamber from the ITM.
e) Gently remove the plug-in gas sensor from the old ITM and install it in the new ITM. Orient the plug in
sensor so that it matches with the female connector pins on the new ITM and press the sensor in firmly to
make proper connection.
f) Thread the Splash Guard Adapter and Sample Chamber onto the ITM until snug, tighten the locking
setscrew and reconnect splashguard.
g) Feed the sensor assembly wires through the ¾” female NPT mounting hole and thread the assembly into the
J-box until tight and the ITM lens faces toward the front access point. Connect the sensor assembly wires
inside J-Box (Refer to Section 2.6, and Figure 8).
h) Perform Set AutoSpan Level, Set Serial ID, Set Range, and then perform a successful AutoZero and
AutoSpan before placing sensor into service.