Model PI-700 0-500ppb

PI-700 0-500ppb Instruction Manual

Rev. 0.0

Page 6 of 53

2.3 Sensor Contaminants and Interference


PID VOC and toxic gas sensors will respond to any gas with an ionization potential <10.6eV. This is not a
selective measurement technique, and hence can be used to measure a wide range of gases.

Some of the most commonly present gasses that potentially cause PID interference are listed in Table 2 Gas
Factor Table
(refer to Section 9). The presence of cross-interference gases in an area does not preclude the
use of this sensor technology, although it is likely that the sensor will experience false high readings should
exposure occur.

Some heavy organic molecules may be polymerized onto, or strongly adhere to, the optical filter of the lamp.
When this occurs, the lamp will require cleaning or replacement.

Relative Response Gas Matrix

Table 2 Gas Factor Table shows the response of the PID sensor to a long list of components. It includes the
compound name, synonyms/abbreviations, and chemical formula. It also lists the 10.6eV Response Factor
(the measure of how strong the signal from the sensor is in reference to Isobutylene gas). Isobutylene gas is
the standard reference used with PID sensors, the lower the Response Factor, the stronger the signal.

2.4 Mounting Installation


The PI-700 sensor assembly is designed to be mounted on a flat surface. The assembly comes on a 13” X 14”
mounting plate and is ready to be mounted using 1/4” Bolts. A ¾”NPT Port is provided on the side of the
Mini Condulet for power and signal wiring. A 1/8” Tube fitting is provided for Sample Port gas entry. The
PI-700 should be vertically oriented so that the sensor points straight down.

AIR

FLOW

SPAN

PGM2

ZERO

PGM1

MODEL

PI-700

detcon inc.

VOC

14"

12.5"

11.5"

13"

3

4

" NPT Port

Ø0.265"

Mounting Holes

Aluminum

Condulet with

Pump

Mini

Condulet
with TPA

0-1LPM

Rotameter

PI-700 0-500ppb
Sensor Assembly

Sample Chamber

Sample

Port

Air Flow

Figure 7 Outline and Mounting Dimensions