beautypg.com

2 set autospan level, 3 set serial id, Set autospan level – Detcon PI-700 0-500ppb User Manual

Page 22: Set serial id

background image

Model PI-700 0-500ppb

PI-700 0-500ppb Instruction Manual

Rev. 0.0

Page 18 of 53


Gain Setting

The menu item appears as: “Gain XX”


Raw Counts

The menu item appears as: “Counts XXXX”


When the status list sequence is complete, the ITM will revert to the “View Sensor Status” text scroll. The
user can either: 1) review list again by executing another 3-4 second hold, 2) move to another menu item by
executing a momentary hold over PGM1 or PGM2, or 3) return to Normal Operation via automatic timeout of
about 15 seconds (the display will scroll “View Sensor Status” 4 times and then return to Normal Operation).

3.5.2 Set AutoSpan Level


Set AutoSpan Level is used to set the span gas concentration level that is being used to calibrate the sensor.
This level is adjustable from 10% to 50% of selected full-scale range. The current setting can be viewed in
View Program Status.

The menu item appears as: “Set AutoSpan Level”.

From the Set AutoSpan Level text scroll, hold the magnet over PGM1 or PGM2 until the “V” prompt appears
and continue to hold the magnet in place for an additional 3-4 seconds (until the display starts to scroll “Set
Level”). The display will switch to “ XX“ (where XX is the current gas level). Swipe the magnet
momentarily over PGM2 to increase or PGM1 to decrease the AutoSpan Level until the correct level is
displayed. When the correct level is achieved, hold the magnet over PGM1 or PGM2 for 3-4 seconds to
accept the new value. The display will scroll “Level Saved”, and revert to “Set AutoSpan Level” text scroll.

Move to another menu item by executing a momentary hold, or return to Normal Operation via automatic
timeout of about 15 seconds (the display will scroll “Set AutoSpan Level” 4 times and then return to Normal
Operation).

3.5.3 Set Serial ID


Detcon Model 700 sensors can be polled serially via RS-485 Modbus™ RTU. Refer to Section 4.0 for details
on using the Modbus™ output feature.

Set Serial ID is used to set the Modbus™ serial ID address. It is adjustable from 01 to 256 in hexadecimal
format (01-FF hex). The current serial ID can be viewed in View Sensor Status using the instruction given in
Section 3.5.1 View Sensor Status.

The menu item appears as: “Set Serial ID”.

From the “Set Serial ID” text scroll, hold the programming magnet over PGM1 or PGM2 until the “V” prompt
appears and continue to hold the magnet in place for an additional 3-4 seconds (until the display starts to scroll
“Set ID”). The display will then switch to “ XX“ (where XX is the current ID address). Swipe the magnet
momentarily over PGM2 to increase or PGM1 to decrease the hexadecimal number until the desired ID is
displayed. Hold the magnet over PGM1 or PGM2 for 3-4 seconds to accept the new value. The display will
scroll “ID Saved”, and revert to “Set Serial ID” text scroll.

Move to another menu item by executing a momentary hold, or, return to Normal Operation via automatic
timeout of about 15 seconds (the display will scroll “Set Serial ID” 5 times and then return to Normal
Operation).