beautypg.com

Figure 23 cell assembly installation, Figure 24 spacer installation, Figure 25 installing filter media – Detcon PI-700 0-500ppb User Manual

Page 36

background image

Model PI-700 0-500ppb

PI-700 0-500ppb Instruction Manual

Rev. 0.0

Page 32 of 53


3. Using fine tipped tweezers, install the cell assembly. Align the pins with the corresponding sockets on the

sensor and push down on the end with the pins. Make sure the cell assembly is flush with the lamp
window.

Figure 24 Cell Assembly installation


4. Place the spacer around the assembly.

Figure 25 Spacer installation


5. Place the filter media over the Cell Assembly centered on the top of the sensor. Make sure the filters are

installed in the correct order. Filter Media #2 first, then Filter Media #1 on top, with the shiny side up.

Figure 26 Installing Filter Media