beautypg.com

Figure 20 polishing the lamp, Figure 21 lamp installation, Figure 22 lamp seating – Detcon PI-700 0-500ppb User Manual

Page 35

background image

Model PI-700 0-500ppb

PI-700 0-500ppb Instruction Manual

Rev. 0.0

Page 31 of 53

of rubbing should be enough in most cases. Another indication of cleaning completeness is that about 1/16

th

of

the pad surface is used during the process.

Figure 21 Polishing the Lamp

Reassembly

1. Install the lamp into the sensor, making sure that the lamp’s metalized pads are aligned with the

corresponding excitation springs inside the lamp cavity

Figure 22 Lamp installation


2. With the end of the clean tweezers, or a clean blade of a screwdriver, press down firmly, being careful not

to scratch the surface of the lamp.