3 normal operation, Normal operation, Figure 11 pi-700 software flowchart – Detcon PI-700 0-500ppb User Manual
Page 16: Software flowchart
Model PI-700 0-500ppb
PI-700 0-500ppb Instruction Manual
Rev. 0.0
Page 12 of 53
Software Flowchart
dec
dec
mA Output = XX.XX
Sensor Life XXX%
Last Cal XX Days
Gas Factor = XX
Zero Offset = X.X
inc
PGM1/2 (M)
PGM2 (10)
Defaults Restored
PGM2 (S)
PGM1/2 (3)
PGM1 (S)
Restore Defaults
Auto Time-Out
##
AutoSpan
AutoSpan @ XX
Range XXX
Serial ID XX
Gas Type
Auto Time-Out
Version X.XX
View Sensor Status
Model Type
PGM1/2 (3)
PGM1/2 (M)
inc
PGM1 (S)
PGM1/2 (3)
Auto Time-Out
PGM1/2 (M)
Set Gas Factor
PGM1/2 (3)
AutoTime-out
PGM2 (S)
PGM1/2 (3)
PGM1/2 (M)
Set AutoSpan Level
##
PGM1 (3)
PGM2 (3)
Normal Operation
PGM1 (3)
PGM2 (10)
AutoZero
inc - Increase
dec - Decrease
X, XX, XXX - numeric values
(S) - Momentary Swipe
(M) - Momentary hold of Magnet during text
scroll until the ">" appears, then release
(3) - 3 second hold from ">" prompt
(10) - 10 second hold from ">" prompt
Auto Time-out - 5 seconds
PGM1 - Program Switch Location #1
PGM2 - Program Switch Location #2
LEGEND:
PGM1/2 (3)
Simulation
Auto Time-Out
Set Range
Auto Time-Out
Signal Output Check
Auto Time-Out
Set Zero Offset
Auto Time-Out
PGM1/2 (M)
PGM1/2 (3)
PGM1/2 (3)
Set Serial ID
PGM1/2 (M)
inc
dec
PGM1 (S)
PGM1/2 (10)
PGM1/2 (M)
PGM1/2 (3)
PGM1/2 (3)
PGM2 (S)
##
PGM1/2 (M)
PGM1 (S)
PGM1/2 (3)
inc
PGM2 (S)
##
dec
dec
inc
PGM2 (S)
PGM1/2 (3)
PGM1 (S)
##
Gain Setting = XXX
Input Volt = XX.XX
Temperature = XXX
Raw Counts = XXX
Figure 11 PI-700 Software Flowchart
3.3 Normal Operation
In normal operation, the ITM Display continuously shows the current sensor reading, which will normally
appear as “ 0 ”. Once every 60 seconds the LED display will flash the sensor’s measurement units and gas
type (i.e. ppm VOC). If the sensor is actively experiencing any diagnostic faults, a “Fault Detected” message
will scroll across the display on the ITM display once every minute instead of the units of measure and the gas
type. At any time, while the sensor is in “Fault Detected” mode, PGM1 or PGM2 can be swiped to prompt the
sensor to display a list of the active faults.
In normal operation, the 4-20mA current output linearity corresponds with the full-scale range. The RS-485
Modbus™ serial output provides the current gas reading and complete fault status on a continuous basis when
polled by the master device.