LumaSense Technologies M330 User Manual

Page 35

background image

M330 Manual

Service 35

Top view of new Thermostat location.

Rear of blackbody

Front of blackbody

3. Move Transformer Primary tap from 230 VAC (H3) to 208 VAC (H2) as shown below if

operating on 220 VAC (only). Power now goes between H1 and H2 as shown below. Do
not change H3 to H2 if operating at 230 VAC.

Rear view of M330 with rear panel removed