3 enclosure, 4 battery back pack, 5 handheld wireless display – LumaSense Technologies BoilerSpection-IM Manual User Manual
Page 28
BoilerSpection IM Manual
Troubleshooting 28
6.3 Enclosure
Symptom
Probable Cause
Corrective Action
Overheating,
internal dust,
image quality loss
Leakage in enclosure caused by material
degradation
Check all gaskets and seals
Leakage in enclosure caused by environmental
damage
Replace gaskets and seals
Leakage in enclosure caused by mechanical
vibration
Check mechanical fittings and
enclosure for loose or out of
place components and repair
Failure to focus
Damage to mechanical parts due to vibrations Check integrity of focusing
rod
6.4 Battery Back Pack
Symptom
Probable Cause
Corrective Action
Loss of image
Connection failures ie. Wireless transmitter
failure
Open back plate and check all
video wiring connections
Power failure to cabinet
Check integrity of input
power cable and connectors
Communication failure due to mechanical
vibrations
Check mechanical fittings and
enclosure for loose or out of
place components and repair
Loss of power
Failure to charge
Check output of charger then
replace as necessary
6.5 Handheld Wireless Display
Symptom
Probable Cause
Corrective Action
Loss of image
Channel not set correctly
Check that the channel is set
correctly on receiver. Only the
CH4 dip switch should be at
on position
Receiver not turned on
Check that the selection bar is
slid to Receiver and the red
light is on
Frequency not selected correctly
Check that the frequency
selector is set at 2 for 2.4GHz
Battery discharged
Recharge unit