BoilerSpection IM Manual

Troubleshooting 27

6 Troubleshooting

6.1 Lens

Symptom

Probable Cause

Corrective Action

Poor image – miss
slag buildup

Lens surface contamination due to dust/dirt in
the boiler environment

Clean lens with q-tip or install
air purge

Lens surface contamination during
installation/maintenance

Clean lens

Image offset caused by excessive heat

Install air cooling

Focus change caused by excessive heat

Excessive heat causes lens component to
become opaque

Contact LumaSense
(refer to section 1.7)

Loss of image

Filters cracked because of thermal cycling

Contact LumaSense
(refer to section 1.7)

Nose tip damaged because of excessive heat

Nose tip damage because of vibration failure

Damage to mechanical/internal fixtures
caused by excessive heat

Cooling shroud damaged caused by
missalignment

6.2 Camera

Symptom

Probable Cause

Corrective Action

Complete Power
Failure

Vibrations/internal failures

Contact LumaSense
(refer to section 1.7)

Loose power connections

Disconnect battery and check
all power wiring connections

Excessive heating

Air cooling needed or add
heat shields if hot air or
radiation from the furnace is
directly heating the failed
component or enclosure

Battery Discharged

Recharge Battery