0 available ranges, conversions and resolutions – Condec UPC5110 User Manual
Page 40

Available Ranges, Conversions and Resolutions
37
6.0
Available Ranges, Conversions and Resolutions
NOTE: Non-standard ranges are available by special order.
Calibrator Conversion Factors:
•
kPa = PSI x 6.89476
•
Bar = PSI x 0.0689476
•
mBar = PSI x 68.9476
•
Torr = PSI x 51.7149
•
mm Hg = PSI x 51.7149
•
in Hg (0˚F) = PSI x 2.036
•
Kg/cm
2
= PSI x 0.070308
•
in H
2
O (60˚F) = PSI x 27.71
•
cm H
2
O = PSI x 70.308
NOTE: Display resolution 0.02% of selected range, unless it is not devisable by 1, 2, or 5.
RANGES
RESOLUTION
+-28.000 in Hg / +-387.00 in H
2
O / 2771.0 in H
2
O / +-96.00 kPa / 100.00 PSI
(Standard)
0.005 / 0.05 / 0.5 / 0.02 / 0.02
+-28.000 in Hg / +-387.00 in H
2
O / 2771.0 in H
2
O / +-724.0 mm Hg / 100.00 PSI
0.005 / 0.05 / 0.5 / 0.1 / 0.02
+-760.0 Torr / +-1.0000 Bar / 7.0000 Bar / +-96.00 kPa / 100.00 PSI
0.1 / 0.0002 / 0.001 / 0.02 / 0.02
+-28.000 in Hg / +-387.00 in H
2
O / 1034.0 mBar / +-100.00 in H
2
O / 15.000 PSI
0.005 / 0.05 / 0.2 / 0.05 / 0.002
+-724.0 mm Hg / +-960.0 cm H
2
O / 7031 cm H
2
O / +-96.00 kPa / 689.0 kPa
0.1 / 0.2 / 1 / 0.02 / 0.1
+-28.000 in Hg / +-387.00 in H
2
O / 775.0 mm Hg / +-96.00 kPa / 15.000 PSI
0.005 / 0.05 / 0.1 / 0.02 / 0.002
+-28.000 in Hg / +-387.00 in H
2
O / 5542 in H
2
O / +-724.0 mm Hg / 200.00 PSI
0.005 / 0.05 / 1 / 0.1 / 0.02
+-28.000 in Hg / +-387.00 in H
2
O / 2771.0 in H
2
O / +-14.000 PSI / 100.00 PSI
0.005 / 0.05 / 0.5 / 0.002 / 0.02
+-28.000 in Hg / +-387.00 in H
2
O / 5542 in H
2
O / +-14.000 PSI / 200.00 PSI
0.005 / 0.05 / 1 / 0.002 / 0.02
Table 6-1. Display Resolutions