14
UPC5100/UPC5110 Operation & Maintenance Manual
4.0
Maintenance and Service
This section outlines the mechanical and basic electrical repair procedures for the UPC5100/UPC5110.
4.1
Troubleshooting
Use Table 4-1 below for information on troubleshooting the UPC5100/UPC5110.
4.2
Maintenance and Service Procedures
The repair procedures cover the major components and sub-assemblies which are critical to the proper
functioning of the calibrators and may need periodic maintenance over the life of the unit.
Only those persons who are formally trained as skilled technicians should attempt to repair these units.
Although some mechanical sub-assemblies could be replaced without venting cylinder it is not
recommended. All safety precautions should be observed due to the presence of electrical
components and high-pressure cylinders. Unit must always be unplugged from power source.
4.2.1
UPC5100/UPC5110 Panel/Chassis - Removal and Installation
Tools required:
Phillips screwdriver
UPC5100 Removal:
1. Verify that power toggle switch is in the OFF position. Loosen and remove the eight screws (PN 14862)
that secure the panel assembly to the enclosure.
2. Lift the panel and chassis by grasping the fill port fitting and vent/vacuum port. Then grasp under the
panel edges. Ensure that the wire harnesses do not catch and snag.
3. Gently set the panel/chassis assembly on a bench top. It can be rested on the panel bottom and pressure
cylinder, with the panel tilted at an angle from its vertical.
Symptom
Problem
Remedy
No lit display
Unit will not energize
Check fuse, check power source, check
power switch
Display slowly decreases over time
Leak in system
Check all compression and pipe fittings
with snoop, bottle of liquid leak gas
detector (PN 64781)
Display does not respond when Vernier
knob is turned
No Vernier control
Readjust isolation valves on Orion;
replace O-ring on Vernier piston
Display increases or decreases when
COARSE (Pressure) or VENT valves are
closed
No Pressure or Vent control
Replace valve seats or O-rings in valves;
check valve needles
Unit will not stay in CAL; display shows
"o" and reads a high value at zero PSIG.
Transducer over-pressurized
Replace transducer
Low battery indicator on display
illuminates when unit is powered
Low or no battery power
Re-charge battery, check power supply
charging voltage
No display when in battery mode after
charging
Battery will not hold charge
Replace battery
Display will not zero
Display will not zero
Perform a ZERO/SPAN calibration
Display shifts
Transducer drifts or possible over
pressure
Replace transducer
Gas escapes when external supply
pressure is bled
Nitrogen cylinder will not remain charged Remove inlet check valve; clean or
replace
Table 4-1. UPC5100/UPC5110 Troubleshooting