background image

14

UPC5100/UPC5110 Operation & Maintenance Manual

4.0

Maintenance and Service

This section outlines the mechanical and basic electrical repair procedures for the UPC5100/UPC5110.

4.1

Troubleshooting

Use Table 4-1 below for information on troubleshooting the UPC5100/UPC5110.

4.2

Maintenance and Service Procedures

The repair procedures cover the major components and sub-assemblies which are critical to the proper
functioning of the calibrators and may need periodic maintenance over the life of the unit.

Only those persons who are formally trained as skilled technicians should attempt to repair these units.
Although some mechanical sub-assemblies could be replaced without venting cylinder it is not
recommended. All safety precautions should be observed due to the presence of electrical
components and high-pressure cylinders. Unit must always be unplugged from power source.

4.2.1

UPC5100/UPC5110 Panel/Chassis - Removal and Installation

Tools required:

Phillips screwdriver

UPC5100 Removal:

1. Verify that power toggle switch is in the OFF position. Loosen and remove the eight screws (PN 14862)

that secure the panel assembly to the enclosure.

2. Lift the panel and chassis by grasping the fill port fitting and vent/vacuum port. Then grasp under the

panel edges. Ensure that the wire harnesses do not catch and snag.

3. Gently set the panel/chassis assembly on a bench top. It can be rested on the panel bottom and pressure

cylinder, with the panel tilted at an angle from its vertical.

Symptom

Problem

Remedy

No lit display

Unit will not energize

Check fuse, check power source, check
power switch

Display slowly decreases over time

Leak in system

Check all compression and pipe fittings
with snoop, bottle of liquid leak gas
detector (PN 64781)

Display does not respond when Vernier
knob is turned

No Vernier control

Readjust isolation valves on Orion;
replace O-ring on Vernier piston

Display increases or decreases when
COARSE (Pressure) or VENT valves are
closed

No Pressure or Vent control

Replace valve seats or O-rings in valves;
check valve needles

Unit will not stay in CAL; display shows
"o" and reads a high value at zero PSIG.

Transducer over-pressurized

Replace transducer

Low battery indicator on display
illuminates when unit is powered

Low or no battery power

Re-charge battery, check power supply
charging voltage

No display when in battery mode after
charging

Battery will not hold charge

Replace battery

Display will not zero

Display will not zero

Perform a ZERO/SPAN calibration

Display shifts

Transducer drifts or possible over
pressure

Replace transducer

Gas escapes when external supply
pressure is bled

Nitrogen cylinder will not remain charged Remove inlet check valve; clean or

replace

Table 4-1. UPC5100/UPC5110 Troubleshooting