3.3

Zero/Span Calibration

Selecting the

ZERO/SPAN

position on the Condec Calibration Module (PN 60109) places the instrument into its

ZERO/SPAN

calibration mode. The display is shown in Figure 3-2.

Figure 3-2. Zero/Span calibration for Gage-Only Units

The unit needs to be calibrated to only one positive and one bi-directional range. All bi-directional ranges are
calibrated over 0 to +14 PSI and 0 to -14 PSI. All the positive ranges from 0 to 100 PSI. Starting with one of the
instrument’s ranges, perform the steps shown in Table 3-1. Perform the following for each step:

1. Adjust input pressure to the appropriate (either 0 or full-scale) value.

2. Perform the action indicated in Table 3-1 when pressure input readings are stable.

NOTE: A vacuum pump will be required to enable calibrating the -14 PSI of the bi-directional ranges.

Step No.

Pressure Input Value

Operator Action Required

Resulting Calibration Standard

Display Indication

Remarks

(For Bi-directional

range)

+/-30"

Hg

+/-400"

H

2

0

+/-100

kPa

1

0 PSI

Press ENTER

button

0.000

0.000

0.000

Note 1 below

2

+14 PSI

Press ENTER button

28.505

387.95

96.52

Notes 2 & 3
below

3

0 PSI

Press ENTER button

0.000

0.000

0.000

Note 1 below

4

-14 PSI

Press ENTER button

-28.50
5

-387.9
5

-96.52

Notes 2 & 3
below

Step No.

(For positive direction

range)

Operator Action Required

100
PSI

2700"
H

2

O

Remarks

1

0 PSI

Press ENTER button

0.00

0.00

Note 1 below

2

100 PSI

Press ENTER button

100.00

2771.0

Notes 2 & 3
below

Table 3-1. Zero and Span Calibration Sequence