3 zero/span calibration – Condec UPC5110 User Manual
Page 13
![background image](/manuals/361802/13/background.png)
10
UPC5100/UPC5110 Operation & Maintenance Manual
3.3
Zero/Span Calibration
Selecting the
ZERO/SPAN
position on the Condec Calibration Module (PN 60109) places the instrument into its
ZERO/SPAN
calibration mode. The display is shown in Figure 3-2.
Figure 3-2. Zero/Span calibration for Gage-Only Units
The unit needs to be calibrated to only one positive and one bi-directional range. All bi-directional ranges are
calibrated over 0 to +14 PSI and 0 to -14 PSI. All the positive ranges from 0 to 100 PSI. Starting with one of the
instrument’s ranges, perform the steps shown in Table 3-1. Perform the following for each step:
1. Adjust input pressure to the appropriate (either 0 or full-scale) value.
2. Perform the action indicated in Table 3-1 when pressure input readings are stable.
NOTE: A vacuum pump will be required to enable calibrating the -14 PSI of the bi-directional ranges.
Step No.
Pressure Input Value
Operator Action Required
Resulting Calibration Standard
Display Indication
Remarks
(For Bi-directional
range)
+/-30"
Hg
+/-400"
H
2
0
+/-100
kPa
1
0 PSI
Press ENTER
button
0.000
0.000
0.000
Note 1 below
2
+14 PSI
Press ENTER button
28.505
387.95
96.52
Notes 2 & 3
below
3
0 PSI
Press ENTER button
0.000
0.000
0.000
Note 1 below
4
-14 PSI
Press ENTER button
-28.50
5
-387.9
5
-96.52
Notes 2 & 3
below
Step No.
(For positive direction
range)
Operator Action Required
100
PSI
2700"
H
2
O
—
Remarks
1
0 PSI
Press ENTER button
0.00
0.00
Note 1 below
2
100 PSI
Press ENTER button
100.00
2771.0
Notes 2 & 3
below
Table 3-1. Zero and Span Calibration Sequence