UPC5100/UPC5110 Operation & Maintenance Manual
4.2.3
ORION-2D Manifold Removal (PN 55286)
Tools required:
Phillips screwdriver
11/32" wrench or nutdriver
.061" hex wrench
adjusting screwdriver (small flat blade)
11/32" open end wrench (thin)
7/16" open end wrench
NOTE:
See Table 4-6 on page 33 and Figure 4-1 on page 34 for additional parts information.
1. Vent any remaining nitrogen from cylinder to atmosphere. Disconnect power cord from power source.
2. Remove front panel from its enclosure as described in Section 4.2.1 on page 14, and place unit on a
bench top.
3. Remove the tubing section from the
VENT
outlet fitting on the ORION-2D, using a 7/16" wrench
4. Remove the flexible tubing going to the transducer port. Also the flexible tubing coming from the
vacuum generator, compressed air port, and the mode select assembly.
NOTE:
Marking flexible tubing ends may be helpful for reinstalling ORION-2D.
5. If the transducer is wired via a connector, remove the connector by turning counter-clockwise. If the
transducer is hard-wired, loosen and remove the four transducer wires (red, white, green, black) from the
terminal block, TB1, on the CPU board, using the small flat-blade screwdriver.
6. Break the wire ties that hold the transducer wires so that the wires are free.
7. Using the 11/32" thin wrench, loosen and carefully remove the transducer from the ORION-2D
manifold.
8. Remove the panel knobs from the
COARSE
(pressure),
VERNIER
and
VENT
valves using the .061" hex
wrench.
NOTE: Earlier manufactured units may have the vacuum generator mounted to the ORION-2D manifold body. If so,
remove the muffler, coupling and long nipple assembly from the vacuum generator.
9. While holding ORION-2D, loosen and remove the two panel screws (PN 60837) from the panel front
that secure the manifold to the panel. Remove the ORION-2D.
4.2.4
ORION-2D Manifold - Valve Seat Removal
Tools required:
A/R solvent (de-natured alcohol)
socket wrench
3/4" socket
needle housing socket (PN 65580)
isolation valve needle housing socket (PN 68509)
hex wrench (.050")
hex wrench (.061")
needle-nose pliers
tube fluorinated Krytox grease (PN 55593)
electric hand drill
No. 43 drill bit
No. 4-40 tap
tap handle
small hammer
NOTE:
See Table 4-6 on page 33 and Figure 4-1 on page 34 for additional parts information.
1. Use a bench vise to secure the manifold by its center portion, with the valve knobs pointing upward.
2. Using the .061" hex wrench, loosen and remove the knob inserts (4) from the pressure and vent valve
stems.
3. Loosen the 3/4" locknuts (1) on the
COARSE
(pressure) and
VENT
valve threaded needle housings (10).
4. Using the needle housing socket (65580) and socket wrench, loosen and remove the needle/housing
assembly (10, 11).
5. To disassemble the isolation valves (inner valve), first remove the valve needle (18) by turning the gear
(6) clockwise.
6. Loosen and remove the valve housings (19) using the isolation valve housing removal socket (68509)
and socket wrench.