5 vacuum measurement sequence – Condec UPC5110 User Manual
Page 10
![background image](/manuals/361802/10/background.png)
Operation
7
Figure 2-3. Pressure or Vacuum Measurement Sequence
NOTE: UPC5100 shown, AC Input (7) and Fill Port (4) are on back side of UPC5110 Rack Mountable Calibrator.
2.5
Vacuum Measurement Sequence
See Figure 2-3 above and proceed as follows:
1. Turn the
MODE SELECT
valve (10) to the
VACUUM
position.
2. Close the
COARSE ADJUSTMENT
valve (2) clockwise and open the
VENT
valve (8), counter-clockwise.
3. Zero the unit by pressing the
ZERO
button (12).
4. If only pressure measurements greater than atmospheric pressure are required continue to step 4.1. If
simultaneous pressure measurements above and below atmospheric pressure, or if a vacuum level higher
than 27” Hg are required, go to step 5.
4.1 Close the
VENT
valve (8) (approximately two turns to its stop) and open the
COARSE
ADJUSTMENT
valve (2) slowly by rotating knob counter-clockwise until the numerical display
begins to move. It will immediately activate the vacuum converter. A vacuum as low as 27” Hg can
be maintained in the system without the use of an external vacuum pump. In general, the pressure
may be changed rapidly until reaching approximately 90% of its desired final value. Closing the
COARSE ADJUSTMENT
valve (2), clockwise will deactivate the vacuum converter and trap the
vacuum in the system.
NOTE: The vacuum may be vented by opening VENT valve counter-clockwise or switching the MODE SELECT to the
PRESSURE position and opening the COARSE ADJUSTMENT valve counter-clockwise.
4.2
Use the
COARSE ADJUSTMENT
(2) and
VENT
valve (8) to obtain a specific pressure reading.
Both provide precise control. As the pressure approaches the desired value, the valve being used
for control should be rotated slowly clockwise to its closed position.
4.3
To obtain exact pressure readings, slowly rotate the
VERNIER
control (13) knob in the direction
required (counter-clockwise to increase vacuum) as indicated by the electronic numerical display.
5. If simultaneous pressure measurements above and below atmospheric pressure are required, or if a
vacuum level higher than 27” Hg are required, connect a vacuum pump to the
VACUUM/VENT
port (15)
as shown in Figure 2-3 on page 7.
6. Open the
VENT
valve (8). Close the
COARSE ADJUSTMENT
valve (2) and apply power to the vacuum
pump and allow it to evacuate the system for several minutes or until the digital display reading reaches
equilibrium near zero PSIA. Press the
ZERO
button to establish a zero reference on the display.