beautypg.com

6 battery operation – Condec UPC5110 User Manual

Page 11

background image

8

UPC5100/UPC5110 Operation & Maintenance Manual

7. With the vacuum pump still running, close the VENT valve (8) and check for system leaks. If there are

none, continue to step 7.1

7.1

To apply pressure, close the

VENT

valve (8) (approximately two turns to its stop) and open the

COARSE ADJUSTMENT

valve (2) (approximately 1/2 turn counter-clockwise until the numerical

display begins to move). In general, the pressure may be changed rapidly until reaching
approximately 90% of its desired final value.

7.2

Use either the

COARSE ADJUSTMENT

(2) or

VENT

valve (8) to obtain a specific pressure

reading. Both provide precise control. As the pressure approaches the desired value, the valve
being used for control should be rotated slowly clockwise to its closed position.

7.3

To obtain exact pressure readings, slowly rotate the

VERNIER

control (13) knob in the direction

required (clockwise to increase pressure) as indicated by the electronic numerical display.

2.6

Battery Operation

When supplied with the battery, the UPC5100/UPC5110 has an internal, rechargeable 12 volt, lead acid battery
which provides a minimum of six hours of completely portable usage before having to be recharged.

An

ON/OFF/BATTERY TEST

switch (18) (Figure 2-3 on page 7) is provided to conserve energy when the

instrument is not in use, and to provide the operator with information as to the status of the battery voltage during
use.

The UPC5100/UPC5110 can be operated and recharged by connecting to a standard AC outlet via the line cord
(supplied). The battery re-charge cycle time is approximately 16 to 20 hours with the

ON/OFF

switch in the

OFF

position. The charging circuit is designed to be left on indefinitely without adversely affecting battery life.

When selected, the momentary action

BATTERY TEST

switch (18) is used to read the actual battery voltage. The

battery voltage reading typically is between 11.5 and 13.0 volts. When the battery voltage reads 11.5 volts, there
are approximately one to two hours of useful operation left and a low battery indicator is illuminated. For LED
display units, a LED in the left center of the display turns on. For LCD display units, five LED segments in the
left of the display window illuminate in a “U” shape. The instrument ceases to function when the battery voltage
is 11.0 volts or less.

NOTE: The battery test should only be performed with the UPC5100/UPC5110 operating at zero PSIG (VENT valve open)
and at the conclusion of the test, the unit’s ZERO button will have to be pushed again to re-zero the instrument.

This manual is related to the following products: