4 linearity and hysteresis calibration – Condec UPC5110 User Manual

Page 14

background image

Calibration

11

NOTES:

1.

If readings are not stable or are not within ±20% of zero, the zero correction can’t be entered.

2.

If readings are not stable or are not within ±5% of 100%, the span correction cannot be entered.

3.

For ease of calibration, do the Linearity/Hysteresis calibration , prior to doing ZERO/SPAN of next range.

3.4

Linearity and Hysteresis Calibration

Install the Condec Calibration Module (PN 60109) and select the

LYN/HYS

position of the rotary switch on the

module. This places the UPC5100/UPC5110 into its linearization/hysteresis calibration mode. The display is
shown in Figure 3-3 below.

NOTE:

The zero/span calibration needs to be performed prior to linearity and hysteresis calibration.

Figure 3-3. Linearity and Hysteresis Calibration

Sequentially perform the thirteen steps described in Table 3-2, for each pressure range being calibrated. Perform
the following for each step:

1. Adjust input pressure to the appropriate value without overshooting the setting. If value is overshot, vent

unit and repeat steps.

2. Perform the action as indicated when the readings are stable, 1 to 2 minutes. If it is taking longer, check

system for leaks. If no leaks are found, the CPU or transducer may be defective.

Step

Input Pressure % of Range

CONDEC Calibration Module

Operator Action Required

UPC5100/UPC5110

Display Status Symbol in

Left-most Digit

Remarks

1

0

Press ZERO switch

Upper Circle

Zero on display

2

10

Press ENTER button

Upper Circle

Notes 1 & 2 below

3

20

Press ENTER button

Upper Circle

Notes 1 & 2 below

4

30

Press ENTER button

Upper Circle

Notes 1 & 2 below

5

40

Press ENTER button

Upper Circle

Notes 1 & 2 below

6

50

Press ENTER button

Upper Circle

Notes 1 & 2 below

7

60

Press ENTER button

Upper Circle

Notes 1 & 2 below

8

70

Press ENTER button

Upper Circle

Notes 1 & 2 below

9

80

Press ENTER button

Upper Circle

Notes 1 & 2 below

10

90

Press ENTER button

Upper Circle

Notes 1 & 2 below

Table 3-2. Linearization and Hysteresis Calibration Sequence