Calibration
11
NOTES:
1.
If readings are not stable or are not within ±20% of zero, the zero correction can’t be entered.
2.
If readings are not stable or are not within ±5% of 100%, the span correction cannot be entered.
3.
For ease of calibration, do the Linearity/Hysteresis calibration , prior to doing ZERO/SPAN of next range.
3.4
Linearity and Hysteresis Calibration
Install the Condec Calibration Module (PN 60109) and select the
LYN/HYS
position of the rotary switch on the
module. This places the UPC5100/UPC5110 into its linearization/hysteresis calibration mode. The display is
shown in Figure 3-3 below.
NOTE:
The zero/span calibration needs to be performed prior to linearity and hysteresis calibration.
Figure 3-3. Linearity and Hysteresis Calibration
Sequentially perform the thirteen steps described in Table 3-2, for each pressure range being calibrated. Perform
the following for each step:
1. Adjust input pressure to the appropriate value without overshooting the setting. If value is overshot, vent
unit and repeat steps.
2. Perform the action as indicated when the readings are stable, 1 to 2 minutes. If it is taking longer, check
system for leaks. If no leaks are found, the CPU or transducer may be defective.
Step
Input Pressure % of Range
CONDEC Calibration Module
Operator Action Required
UPC5100/UPC5110
Display Status Symbol in
Left-most Digit
Remarks
1
0
Press ZERO switch
Upper Circle
Zero on display
2
10
Press ENTER button
Upper Circle
Notes 1 & 2 below
3
20
Press ENTER button
Upper Circle
Notes 1 & 2 below
4
30
Press ENTER button
Upper Circle
Notes 1 & 2 below
5
40
Press ENTER button
Upper Circle
Notes 1 & 2 below
6
50
Press ENTER button
Upper Circle
Notes 1 & 2 below
7
60
Press ENTER button
Upper Circle
Notes 1 & 2 below
8
70
Press ENTER button
Upper Circle
Notes 1 & 2 below
9
80
Press ENTER button
Upper Circle
Notes 1 & 2 below
10
90
Press ENTER button
Upper Circle
Notes 1 & 2 below
Table 3-2. Linearization and Hysteresis Calibration Sequence