Calibration

13

2. Press the

ENTER

button on the module. The display reads

0.00

.

3. Set the current generator for 20 mA output. Press the

ENTER

button on the module. The display should

read

100.000

.

4. Turn the

DISPLAY SELECT

switch (16) to the

CURRENT

position. Display will read 20.000.

5. Disconnect the current generator.

NOTE:

If the display reading is off, set the Current Generator to 0, and press the ENTER button on the Condec

Calibration Module. Set the Current Generator for 20 mA output. The display will read 20.000. If the display reading is
off, press the ENTER button on the module. If the display reading is not 20.000, CPU is faulty and requires servicing.

3.7

Permanent Data Storage

After completing the above calibration procedures, the new data that has been entered into the computer must be
permanently stored. The sequence to do this is as follows:

1. Select the

DATA RECALL

position of the rotary switch on the Condec Calibration Module (PN 60109).

2. Press the

STORE

button on the module.

3. When the data is accepted, the four-digit number on the display indicates

1

020

for as long as the

STORE

button is pressed.

3.8

Normal Mode Test

After completing the above calibration procedures, you must perform a normal mode test.

1. Set the Condec Calibration Module to the

NORMAL MODE

position.

2.

DISPLAY SELECT

switch should still be in the

CURRENT

position. Display will read

20.000

.

3. Turn the

DISPLAY SELECT

switch to the

VOLTAGE

position. Display will read

100.00

.

4. The pneumatic portion of the calibration is now complete and the pressure standard and the module can

now be disconnected.