beautypg.com

Mv stores setting data on a ram disk, Ml reads set data from the ram disk, Me deletes a file on a ram disk – Yokogawa DR240 User Manual

Page 72

background image

6-18

IM DR231-11E

LOGIC

Data are stored each time the
specified event is generated.

p4

Data length
(10, 20, 40, 50, 100, 200, 300, 400, 500, 1k,
2k, 3k, 4k, 5k, 10k, 20k, 30k, 40k, 50k)

p5

Pretrigger

Example

After generating a trigger signal, store only a file of
data length of 3K of the measured
data. Set the retention interval at 2 min, the
pretrigger to 10%.

MWTRIGER, TEST, 2 min, 3K, 10

Comments

• This command is effective only with a floppy disk

drive function.

• If p1 is in DIRECT mode, the pretrigger setting

for p5 will be invalid.

• If p1 is in the TRIGGER or REPEAT mode, the

event/action functions need to set MEMRY:WR
_TRIG for its action.

• If LOGIC3 is set for p3, the event/action functions

need to set MEMRY:DATA_WR for their action.

• AUX, CON, PRN, NUL, or CLOCK cannot be

used in a file name. In addition, no space can be
used for the first character.

• When the pretrigger is set, the data ranges that

have been set before generating a trigger signal
are stored. Pretrigger setting is done in a
percentage of the data length. If 0 is set, the data
before generating a trigger signal cannot be stored.

MR

Reads measured data that have been
stored on a RAM disk.

Mode

Operation mode

Setting

MRp1, p2, p3

p1

Read mode

DIRECT

Reads measured data
immediately after sending a
MR command.

TRIGGER Reads measured data after

generating a trigger signal.

p2

File name (up to eight characters)

p3

Selection to begin reading data (YES/NO)

Example

Read the measured data of a file name TEST
immediately after sending an MR

command.

MRDIRECT, TEST, YES

Comments

• This command is effective only with an optional

floppy disk drive function.

• If the read mode is in TRIGGER, the setting of p3

will be invalid.

• If p1 is in the TRIGGER mode, the event/action

functions need to set MEMRY:RD_TRIG for their
action.

MV

Stores setting data on a RAM disk.

Mode

Operation mode

Setting

MVp1

p1

File name (up to eight characters)

Example

Store the setting data with the file name SET1.

MVSET1

Comments

• This command is effective only with a floppy disk

drive function.

• Data set in the setup mode cannot be stored. To

store the data seting in the setup mode, use a YV
command.

ML

Reads set data from the RAM disk.

Mode

Operation mode

Setting

MLp1, p2

p1

Method of reading the set data:

DIRECT

Starts reading immediately
after sending an ML command.

TRIG1 to 3 Starts reading at the same time

as an event occurrence in
event/action functions.

p2

File name (up to eight characters)

Example

Read the set data with the file name SET1
immediately.

MLDIRECT, SET1

Comments

• This command is effective only with an optional

floppy disk drive function.

• Setting data in the setup mode cannot be read. To

read the setting data in setup mode, use YL
command.

• If any of TRIG1 to 3 is set to p1, the event/action

functions need to set MEMRY:LD_TRG1 to 3 for
their action.

• This command is not executable during

computation.

ME

Deletes a file on a RAM disk.

Mode

Operation mode

Setting

MEp1, p2

p1

Type of file

DATAmeasured/computed data
PANEL

Setting data

p2

File name (up to eight characters)

Example

Delete a file with measured data (TEST)

MEADATA, TEST

Comments

• This command is effective only with an optional

floppy disk drive function.

• A setting data file in the setup mode cannot be

deleted. To delete it, use YE command.

MY

Makes a copy of files between the
RAM and floppy disks.

Mode

Operation mode

Setting

MYp1, p2, p3, p4, p5, p6, p7, p8

p1

Copy destination/copy source

TO_FDD

Copies a file from the RAM
disk to the floppy
disk.

FROM_FDD Copies a file from the floppy

disk to the RAM disk.

p2

Type of file

DATAMeasured/computed data
PANEL

Seting data

p3

File name (up to eight characters)

p4

ASCII conversion on/off

p5

First channel number for ASCII conversion

p6

Last channel number for ASCII conversion

p7

First data number for ASCII conversion

p8

Last data number for ASCII conversion

Example

Make an ASCII conversion of data numbers 1 to 100
from the input channels 005 to
010 in the file TEST on the RAM disk and make a
copy of them on a floppy disk.

MYTO _FDD, DATA, TEST, ON, 005, 010, 1, 100

Comments

• This command is effective only with an optional

floppy disk drive function.

• Parameters from p4 to p8 are effective only when

p2 is set to DATA.

• Parameters from p5 to p8 are effective only when

p4 is on.

6.6 Control and Execution Command

This manual is related to the following products: