Sn sets the scaling unit, Xq sets the filter, Xv sets the measurement period – Yokogawa DR240 User Manual
Page 57: Xi sets the a/d integration time

6-3
IM DR231-11E
Commands
6
• For the measurement item p6, refer to the list of
measurement items on page 5-9 and 5-10.
• There are items that can not be set on the
measurement item, p6, depending on the wiring
method or the settings on other channels. Set
within the group indicated on page 5-10.
• Set the span values, p7 and p8, within the
measurement range indicated on page 5-8.
• Input values within 6 digits disregarding the
decimal point for p7 and p8. The decimal point
position is fixed to the decimal point position of
the setting range of page 5-8.
• For p9, p10, and p11, either set all the three
parameters or omit all three.
• The measurement range cannot be changed while
executing a report or computation, or when a file
is written.
Setting the strain input scaling(DR232/DR242)
Setting
SRp1, p2, p3, p4, p5, p6, p7, p8, p9
p1
Channel number (001 to 560)
p2
SCL
p3
STRAIN
p4
Measurement range
2k/20k/200k
p5
Left value of span
p6
Right value of span
p7
Left value of scale
(-30000 to 30000)
p8
Right value of scale
(-30000 to 30000)
p9
Decimal point position of scale (0 to 4)
Example
Measure using the measurement range of 2k for
subunit 0 and channel 01, and convert 0
µε
to 0.00
and 1000
µε
to 100.00.
SR001, SCL, STRAIN, 2k, 0, 1000, 000, 10000, 2
Comments
• Even when the settings of p4 are same parameters,
the measurement range of the strain varies
depending on the gauge method. For details, refer
to page 5-8.
• Set spans p5 and p6 within the measurement range
indicated in page 5-8.
• Input values up to 6 digits disregarding the
decimal point for p5 and p6. The decimal point
position is fixed to the decimal point position of
the setting range of page 5-8.
• For p7, p9, and p10, either set all the three
parameters or omit all three.
• The measurement range cannot be changed while
executing a report or computation, or when a file
is written.
SN
Sets the scaling unit.
Mode
Operation mode
Setting
SNp1, p2
p1
Channel number (001 to 560)
Computation channel number (A01 to A60)
p2
Unit character string (up to 6 characters)
Example
Set the scaling unit for channel 01 of subunit 0 to
ABCDEF.
SN001, ABCDEF
Comments
• For a character string that can be used as units, see
the characters in Section 5.6, “ASCII Code
Table.” However, note that a semi-colon “;”
cannot be used.
• The optional computation channel number for a
stand-alone model is A01 to A30.
XQ
Sets the filter.
Mode
Setup mode
Setting
XQp1
p1
ON
Filter on
OFF
Filter off
Example
Insert a filter in the input.
XQON
XV
Sets the measurement period.
Mode
Setup mode
Setting
XVp1
p1
Measurement interval (0.5, 1, 2, 3, 4, 5, 6,
10, 12, 15, 20, 30, or 60)
Example
Set the measurement period to 10 seconds.
XV10
Comments
• 0.5s and 1s are only for DR232 and DR242.
• The shortest measurement period that can be set
differs depending on the A/D integration time and
the filter on/off setting as follows.
Integration time: Equivalent to 50/60 Hz
Input module
Filter off
Filter on
10CH
0.5 s*
3 s
20CH
2 s
4 s
30CH
2 s
4 s
Integration time: 100 ms
Input module
Filter off
Filter on
10CH
4 s
12 s
20CH
5 s
15 s
30CH
6 s
20 s
*: 2s if the power monitor module is installed
XI
Sets the A/D integration time.
Mode
Setup mode
Setting
XIp1, p2
p1
Subunit (0 to 5)
p2
Integration time (AUTO, 50 Hz, 60 Hz, 100
ms)
Example
Set the A/D integration time in subunit 0 to 50 Hz.
XI0, 50Hz
6.1 Setting the Input