2 specifications, Communication specifications, When supporting rs-232-c commands – Yokogawa DR240 User Manual
Page 32: When reading in instantaneous value data, When displaying communication conditions, Protocols

4-3
IM DR231-11E
Overview and Specifications of Ethernet Interface
4
4.2 Specifications
Communication Specifications
Transmission specifications
: 10BASE-T (CSMA/CD, 10Mbps, Baseband)
Electrical/Mechanical specifications: Conforms to IEEE802.3 (Frames are not supported.)
Protocols
: TCP, IP, UDP, ARP, ICMP
When supporting RS-232-C commands
Communication format
: TCP/IP
Keepalive
: Turn ON/OFF using dip switch 3. When turned ON, it de-
tects timeouts and disconnects communication
Command data
: ASCII
Response data
: ASCII, BINARY
Receive buffer length
: 200 bytes
Maximum number of connections
: 1
Port No.
: 34150
When reading in instantaneous value data
Communication format
: TCP/IP
Keepalive
: Turn ON/OFF using dip switch 3. When turned ON, it de-
tects timeouts and disconnects communication
Command data
: ASCII
Response data
: ASCII, BINARY
Receive buffer length
: 200 bytes
Maximum number of connections
: 4
Port No.
: 34151
When displaying communication conditions
Communication format
: TCP/IP
Command data
: ASCII
Response data
: ASCII
Receive buffer length
: 200 bytes
Maximum number of connections
: 1
Port No.
: 34159
Protocols
Application layer
Transport layer
Network layer
Data link layer
Physical layer
DARWIN services
TCP UDP
IP ICMP ARP
Ethernet (10BASE-T)
IP: Internet Protocol
TCP: Transmission Control Protocol
UDP: User Datagram Protocol
ICM: FInternet Control Message Protocol
ARP: Address Resolution Protocol