60 mV, 6 V
Example
Calibrate the 20-mV range on the module in slot 1 of
subunit 0.
XZ0, 1, CAL/EXEC, 20 mV
Correcting the calibration value manually
(universal module)
Setting
XZp1, p2, p3, ..... , p25
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 20-mV range
p5
Span calibration value for 20-mV range
p6
Zero calibration value for 60-mV range
p7
Span calibration value for 60-mV range
p8
Zero calibration value for 200-mV range
p9
Span calibration value for 200-mV range
p10
Zero calibration value for 2-V range
p11
Span calibration value for 2-V range
p12
Zero calibration value for 6-V range
p13
Span calibration value for 6-V range
p14
Zero calibration value for 20-V range
p15
Span calibration value for 20-V range
p16
Zero calibration value for 50-V range
p17
Span calibration value for 50-V range
p18
Zero calibration value for Pt:1 mA
p19
Span calibration value for Pt:1 mA
p20
Zero calibration value for Pt:2 mA
p21
Span calibration value for Pt:2 mA
p22
Span calibration value for Pt:1 mA-H
p23
Span calibration value for Pt:2 mA-H
p24
Zero calibration value for Cu:2 mA
p25
Span calibration value for Cu:2 mA
Correcting the calibration value manually (DCV/TC/
DI input module)
Setting
XZp1, p2, ..... , p17
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 20-mV range
p5
Span calibration value for 20-mV range
p6
Zero calibration value for 60-mV range
p7
Span calibration value for 60-mV range
p8
Zero calibration value for 200-mV range
p9
Span calibration value for 200-mV range
p10
Zero calibration value for 2-V range
p11
Span calibration value for 2-V range
p12
Zero calibration value for 6-V range
p13
Span calibration value for 6-V range
p14
Zero calibration value for 20-V range
p15
Span calibration value for 20-V range
p16
Zero calibration value for 50-V range
p17
Span calibration value for 50-V range
Correcting the calibration value manually (mA
input module)
Setting
XZp1, p2, p3, p4, p5
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 20-mA range
p5
Span calibration value for 20-mA range
Correcting the calibration value manually (Strain
input module)
Setting
XZp1, p2, p3, ....., p9
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 2k range
p5
Span calibration value for 2k range
p6
Zero calibration value for 20k range
p7
Span calibration value for 20k range
p8
Zero calibration value for 200k range
p9
Span calibration value for 200k range
Correcting the calibration value manually (Digital
input module)
Setting
XZp1, p2, ..... , p7
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
DISPLAY
p4
Zero calibration value for 60-mV range
p5
Span calibration value for 60-mV range
p6
Zero calibration value for 6-V range
p7
Span calibration value for 6-V range
Example
Modify the span calibration value for the 6V range of
the module in slot 1 of subunit 0 to 32000.
XZ0, 1, DISPLAY, , , , 32000
Storage of A/D calibration data in a module (A/D
calibration END processing)
Setting
XZp1, p2, p3, p4
p1
Unit number (0 to 5)
p2
Slot number (0 to 5)
p3
END
p4
Storage selection (ABORT, STORE)
Example
Store the calibrated values for the module in slot 1 of
subunit 0.
XZ0, 1, END, STORE
Comments
• You must execute STORE using A/D calibration
end processing each time after making an A/D
calibration or adjustment for one module. If A/D
calibration end processing (STORE) is not
executed, the calibrated data will not take effect.
• Since it takes an indefinite time to execute the XZ
command, return an ACK after the processing
ends.
• On the controller side, execute the following
processing after receiving an ACK after
transmitting the XZ command. The format of the
ACK status is as follows:
E0Cr+Lf
The received command was normally
processed.
E1Cr+Lf
There is an error in the received
command.
• If A/D calibration data are adjusted, inputting a
numeric value which is out of the setting range
disables the modules to be recognized. For
details, see the separate manual (IM DR231-01E
or IM DR232-01E).
6.5 Other Settings