beautypg.com

Appendix 2, Temperature compensation matrix, Appendix 2, temperature compensation matrix – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 59

background image

51

IM 12D06D05-01E

A

APPENDICES

The matrix can be cleared before entering new

values. Next new matrix values can be entered

as described above. The EXAxt can interpolate

the matrix. During this process it will check if

the matrix is completely ascending/descend-

ing. This is necessary as otherwise the lookup

function can give two results for one tempera-

ture. If an error is found, the EXAxt will specify

the location of the error as shown in the user

interface screen “user defined 1/2”.

The backspace key should be used for delet-

ing an individual matrix value.

An empty value is shown as

APPENDIX 2

Temperature compensation matrix.

1. A minimum number of values is required to make interpolation possible.

The highlighted values marked as

are mandatory to enter.

Concent

Tref

T1

....

Tx

....

T10

Sol1

C1

S1Tr

S1T1

S1T10

....
Solx

SxTx

....
Sol10

C10

S10Tr

S10T1

S10T10

2. Tref (reference temperature) is defined in the Temperature Compensation menu. If Tref is

between T1 and T10 then the value of Tref needs to be entered as Tx (T2....T9).

Concent

Tref

T1

....

Tx

.... T10

Sol1

C1

S1Tr

S1T1

S1T10

....
Solx
....
Sol10

C10

S10Tr

S10T1

S10T10

3. For every SxTx that is entered the following values become mandatory to enter:

Cx, SxTr, SxT1, SxT10 and Tx.

Concent

Tref

T1

....

Tx

.... T10

Sol1

C1

S1Tr

S1T1

S1T10

....
Solx

Cx

SxTr

SxT1

SxTx

SxT10

....
Sol10

C10

S10Tr

S10T1

S10T10