beautypg.com

General specifications of exaxt isc450g – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual

Page 10

background image

2

IM 12D06D05-01E

2. GENERAL SPECIFICATIONS OF EXAxt ISC450G

A) Input specifications : Compatible with the Yokogawa inductive conductivity ISC40 series with

integrated temperature sensor: NTC30k or Pt1000

B) Input range

Conductivity

: 0 to 1999 mS/cm at 25 °C (77 °F) reference temperature

Minimum

: 1 µS/cm (at process temperature)

Maximum

: 2 S/cm (at process temperature)

Temperature

: -20 to +140 ºC (0 to 280 ºF)

Cable length

: max. 60 metres (200 feet)

10 metres (35 feet) fixed sensor cable + 50 metres (165 feet)

WF10 extension cable. Influence of cable can be adjusted by doing

an AIR CAL with the cable connected to a dry cell

C) Accuracy (under reference conditions):

Conductivity

: ≤ 0.5 % of reading ± 1.0 µS/cm

Temperature

: ≤ 0.3ºC (0.6ºF)

Temp.compensation: ≤1 % for NaCl, ≤ 3 % for Matrix

Step response

: ≤ 4 seconds for 90 % (for a 2 decade step)

Note on performance specifications

The following tolerance is added to above performance.

mA output tolerance: ±0.02 mA of “4-20 mA“

D) Transmission signals

General

: Two isolated outputs of 4-20 mA. DC with common negative. Maximum

load 600Ω. Bi-directional HART

®

digital communication, superimposed

on mA1 (4-20 mA) signal.

Output function

: Linear or 21-step table for Conductivity, Concentration or Temperature.

Control function

: PID control.

Burn out function : Burn up (21.0 mA) or burn down (3.6 mA) to signal failure. acc.

NAMUR NE43.

Parameters

: Adjustable damping

: Expire time

Hold

: The mA-outputs are frozen to the last/fixed value during calibration/

commissioning

Cond. range

: min span 10 µs/cm≥10% of high valve (20 mA)

max span 1999 mS/cm

E) Contact outputs

General

: Four SPDT relay contacts with display indicators

Switch capacity

: Maximum values 100 VA, 250 VAC, 5 Amps. (*)

Maximum values 50 Watts, 250 VDC, 5 Amps. (*)

Status

: High/Low process alarms, selected from conductivity, concentration or

temperature. Configurable delay time and hysteresis

: PID duty cycle or pulsed frequency control

: FAIL alarm

Control function

: On / Off

: Adjustable damping

: Expire time