7-1. Periodic maintenance

The converter requires very little periodic main-

tenance, except to make sure the front window

is kept clean in order to permit a clear view of

the display and allow proper operation of the

touchscreen. If the window becomes soiled,

clean it using a soft damp cloth or soft tissue.

To deal with more stubborn stains, a neutral

detergent may be used.

When you must open the front cover and/or

glands, make sure that the seals are clean and

correctly fitted when the unit is re-assembled

in order to maintain the housing’s weatherproof

integrity against water and water vapor.

Note! Never use harsh chemicals or solvents.

In the event that the window does

become heavily stained or scratched,

refer to the parts list (Chapter 10) for

replacement part numbers.

Battery

The EXAxt converter contains a logbook

feature that uses a clock to provide the timings.

The instrument contains a lithium cell (battery)

to support the clock function when the power

is switched off. The cell has an expected work-

ing life of 10 years. Should this cell need to

be replaced, contact your nearest Yokogawa

service center.

Fuse

There is a circuit board mounted fuse protect-

ing the instrument. If you suspect that this

needs to be replaced, contact your nearest

Yokogawa service center.

7-2. Periodic maintenance of the sensor

Note! Maintenance advice listed here is

intentionally general in nature. Sensor

maintenance is highly application

specific.

In general conductivity measurements do not

need much periodic maintenance. If the EXAxt

indicates an error in the measurement or in the

calibration, some action may be needed (ref.

chapter 8 troubleshooting).

7-3. Cleaning methods

1. For normal applications hot water with

domestic washing-up liquid added will be

effective.

2. For lime, hydroxides, etc., a 5 ...10%

solution of hydrochloric acid is

recommended.

3. Organic contaminants (oils, fats, etc.) can be

easily removed with acetone.

4. For algae, bacteria or moulds, use a solution

of domestic bleach (hypochlorite).

* Never use hydrochloric acid and bleaching

liquid simultaneously. The release of the very

poisonous chlorine gas will result.

7-4. LCD adjustment

Contrast adjustment

During the life of the analyzer the contrast of

the display may fade. The contrast can be ad-

justed using the potentiometer on the backside

of the LCD board. This adjustment must be

done only by Yokogawa’s service personnel.

The position is shown on the picture below. For

units manufactured after July 2007, the poten-

tiometer is placed behind the little hole in the

LCD bracket as shown in Figure 3-4 on page 7.

For units manufactured between April 2006

and April 2007, the potentiometer is located as

shown below.

7. MAINTENANCE