3-14. hart device revision, 3-15. logbook, 3-16. trouble shooting – Yokogawa ISC450 4-Wire Analyzer for Inductive Conductivity User Manual
Page 26: 4. information function, 5. setup-calibration & commissioning, 6. secondary- primary value display switch
![background image](/manuals/360183/26/background.png)
18
IM 12D06D05-01E
4-3-14. HART Device revision
Sometimes the firmware of a device is
updated in a way that the communication file
(HART DD) need revision too. In this case the
revision level is increased by one. The revision
level of the HART DD must match the revision
level of the Firmware. The revision level is
expressed by the first two characters of the
filename. The following files should be used
when the HART Device revision level is 2.
(
0201.aot, 0201.fms, 0201.imp, 0201.sym)
4-3-15. Logbook
The EXAxt contains several logbooks to store
historical information on events, changed
settings and calibrations. The logbooks have
been categorized to simplify the retrieval of this
information.
Calibration will give information of previous
calibrations. This logbook is useful as one now
can
1) Monitor the sensor performance over time.
2) Monitor the sensor(s) lifetime.
Sensor will give historical information on
parameter settings concerning the sensor(s).
The events logged in this logbook are user
definable. This is done in:
Commissioning >> Configure Logbook >>
Sensor Logbook.
Settings wil give all history information on pa-
rameter settings concerning the analog outputs
(mA1/mA2) and contact (S1…S4). This logbook
is useful to trace back differences in perform-
ance due to changed settings. The events
logged in this logbook are user definable. This
is done in:
Commissioning >> Configure Logbook >>
Settings Logbook – mA and/or Settings
Logbook – contact
mA1/mA2 shows all (dynamic) events concern-
ing the analog outputs
S1/S2/S3/S4 shows all (dynamic) events con-
cerning the contacts.
Each HMI screen can contain up to 5 events.
As each logbook can contain 50 events in total,
one can access previous events by selecting
another page 1 to 10.
4-3-16. Trouble shooting
If you contact the local sales/ service organiza-
tion the serial number and software revision is
necessary information. Without that information
it is impossible to help you. It is also very useful
to report all the information that you find on the
zoom-in display.
4-4. Information function
In this field an information sign , a warning
sign or a fail sign can appear. Pushing
this button, the user gets detailed information
about the status of the sensor or the instrument
if applicable.
See troubleshooting (chapter 8) for further
details.
4-5. Setup-Calibration & commissioning
By pressing the setup key, you get access
to the operating system of the converter based
on menus and submenus.
Browse through the list using the key
till you find the required menu and press the
key to enter this menu.
It is also possible to press on the or
sym-
bol found beside the menu item.
4-6. Secondary- primary value display switch
Pressing on this text block automatically
switches the secondary value to the main
display (Large font size).
25.0