Yokogawa EJX930A User Manual
Page 39
![background image](/manuals/360056/39/background.png)
<5. Modbus Communication>
5-8
IM 01C25R05-01EN
Address Label used in
DTM
Address Type Access Type
Char/
Date
Read/
Write Enum list
Default
Value
*1
Unit
Valid
Range
IM
Chapter
1132-
1139
Extra No.
Holding
Register
USIGN32×4
ASCII
*2
R
-
spaces
-
-
1140
Final asmbly
num
Holding
Register
USIGN24
-
W
-
0x00000000
-
-
1142-
1149
PT100 Serial
No.
Holding
Register
USIGN32×4
ASCII
*2
W
-
spaces
-
-
1150
Country
Holding
Register
USIGN16
ASCII
W
-
“JP”
-
-
1151
Auto recover
Holding
Register
USIGN8
-
W
Table 33
0x0001
-
0 to 1
1152
Isoltr matl
Holding
Register
USIGN8
-
W
Table 34
-
-
With in
enum list
1153
Fill fluid
Holding
Register
USIGN8
-
W
Table 35
-
-
With in
enum list
1154
Gasket matl
Holding
Register
USIGN8
-
W
Table 36
-
-
With in
enum list
1155
Process Conn
matl
Holding
Register
USIGN8
-
W
Table 37
-
-
With in
enum list
1156
Drain vent matl Holding
Register
USIGN8
-
W
Table 38
-
-
With in
enum list
1157
Process Conn
type
Holding
Register
USIGN8
-
W
Table 39
-
-
With in
enum list
1158
RS isoltr matl
Holding
Register
USIGN8
-
W
Table 40
-
-
With in
enum list
1159
Process Conn
size
Holding
Register
USIGN8
-
W
Table 41
-
-
With in
enum list
1160
Num of RS
Holding
Register
USIGN8
-
W
Table 42
-
-
With in
enum list
1161
RS fill fluid
Holding
Register
USIGN8
-
W
Table 43
-
-
With in
enum list
1162
RS type
Holding
Register
USIGN8
-
W
Table 44
-
-
With in
enum list
1351
Slave Address
Holding
Register
USIGN8
-
W
-
0x0001
-
1 to 247
3.2
1352
Stop Bit
Holding
Register
USIGN8
-
W
Table 55
0x0000
-
0 to 1
3.2
1353
Parity
Holding
Register
USIGN8
-
W
Table 56
0x0000
-
0 to 2
3.2
1354
Turnaround
Delay Time
Holding
Register
USIGN16
-
W
-
0x000A
-
0x000A to
0x00C8
3.2
1401
Pres Unit
Holding
Register
USIGN8
-
W
Table 60
0x000C
-
With in
enum list
3.3.4
1402
SP Unit
Holding
Register
USIGN8
-
W
Table 61
0x00ED
-
With in
enum list
3.3.4
1403
ET Unit
Holding
Register
USIGN8
-
W
Table 62
0x0020
-
With in
enum list
3.3.4
1421
SP H/L Select
Holding
Register
USIGN8
-
W
Table 70
0x0000
-
0 to 1
3.4.1
1422
A/G Select
Holding
Register
USIGN8
-
W
Table 71
0x0001
-
0 to 1
3.4.1
1423
Atm. Pres Value Holding
Register
FLOAT
-
W
-
0.101325
MPa
-2.5 to
27.5
1451
Pres URV
Holding
Register
FLOAT
-
W
-
100.0
kPa
-110 to
110
3.3.3
1453
Pres LRV
Holding
Register
FLOAT
-
W
-
0.0
kPa
-110 to
110
3.3.3
1455
SP URV
Holding
Register
FLOAT
-
W
-
16.0
MPa
-2.5 to
27.5
3.3.3
1457
SP LRV
Holding
Register
FLOAT
-
W
-
0.0
MPa
-2.5 to
27.5
3.3.3
1459
ET URV
Holding
Register
FLOAT
-
W
-
850.0
°C
-210 to
860
3.3.3
*1:
This value is factory default of M range measurement span without order specification.
*2: