beautypg.com

Teledyne LeCroy QPHY-SAS3 User Manual

Page 18

background image

Drive Letter for Matlab

This variable is used for allows the user to specify the location of the waveforms to be used when running
the EyeOpening script remotely over a network. The default value for this variable is D:\.

Path to saved Waveforms for offline test

This variable allows the user to specify the location of the waveforms to be used when running QPHY-
SAS3 in Use Saved Data mode. The default value for this variable is D:\Waveforms\SAS3\.

Save Individual Runs

When this value is set to "Yes", acquired waveforms will be saved in a separate folder for each time the
test is run for example:
D:\Waveforms\SAS3\[Device Under Test]\Run1

When this value is set to "No" waveforms will be overwritten on every run and saved in:
D:\Waveforms\SAS3\[Device Under Test]

This setting has no effect when testing is on saved waveforms.

The default value for this variable is No.

Stop On Test

When set to Yes, the script stops after each test allowing you to view the results. Any new acquisition
done may cause the script to produce unexpected results. The default value for this variable is No.

Path to stressor file for WDP script

This variable defines the path the stressor file which is used in

the SASWDP script. The default value for

this variable is D:\Waveforms\SAS3.

Test Mode

This variable allows the user to choose to run QPHY-SAS3 on newly acquired data or on previously saved
data. The default value for this variable is Acquire New Data.

Tx Negative Source

The variable allows the user to select the input source used for Tx negative signal. The default value for
this variable is C3.

Tx Positive Source

The variable allows the user to select the input source used for Tx positive signal. The default value for
this variable is C2.

Use Sierra to Control Test Pattern

This variable allows the user to specify if a Sierra, Teledyne LeCroy Protocol Analyzer, is being used to
control the test patterns. This is used to generate the proper connection diagram. The default value for
this variable is No.

Use De-embedding

This variable allows the user to specify if the de-embedding feature in SDA3 will be used. The setup file
will in use is defined by the VirtualProbe Path variable. The default value for this variable is No.

VirtualProbe Path

If the user is using the de-embedding feature within SDA3, this variable allows the user to specify if the
path for the VirtualProbe setup file. The default value for this variable is
D:\Applications\VirtualProbe\SAS3\VirtualProbe.lss.

18

922545 Rev A