Battery maintenance – SKC Limited Haz-Dust IV User Manual
Page 68

Battery Maintenance
Introduction
The battery pack is a 6.0 V NiMH rechargeable battery that can hold a charge
for up to 8 hours. It is important to check the battery periodically and
recharge when necessary.
Checking the
Battery
The battery status can be checked using the menu options on the Haz-Dust
IV. Use the following menu options to check the battery.
Step Action
1
Select Special Functions from the Main Menu.
2
Select Systems Options.
3
Select Extended Options.
4
Select Battery Status.
Result: The Battery Level Screen displays the charging level of
the unit’s battery in VDC.
5
If the charge level is...
Then...
6.2 VDC or higher
The battery is fully charged.
6.2 VDC or lower
The battery must be recharged.
See instructions on page 5-9.
Continued on next page
5 - 8