Inhalable dust particulates – SKC Limited Haz-Dust IV User Manual
Page 30

Inhalable Dust Particulates,
Continued
7
Follow the steps in the table below to prepare the IOM sampling
adapter for use with the Haz-Dust IV.
Step
Action
1
Insert rubber washer into IOM cassette front.
Place into sensor and screw the IOM front plate
over it into the sensor.
2
Attach the cassette grid to the filter ring.
3
Push end cover over cassette grid and filter ring.
See figure 3.5 for visual aid.
8
Attach the air intake tubing to the end cover.
9
Turn to page 3-7 and follow the instructions to Auto-Zero the Haz-
Dust IV.
Figure 3-5. Disassembled IOM Sampling Adapter. SKC catalog number 225-70.
3 - 6