Specifications – SKC Limited Haz-Dust IV User Manual
Page 15

Specifications
Introduction
The Haz-Dust IV meets the following specifications.
SPECIFICATION RANGE
Calibration
NIOSH 0600 with SAE Test Dust
Accuracy + 10%
Precision 0.02
mg/m
3
Sensing range
0.01 to 200 mg/m
3
Particle size range
0.1 to 10
µm Respirable
0.1 to 50
µm Thoracic
0.1 to 100
µm Inhalable (IOM)
Recording time
1 second, 1 minute and 10 minute averages
Flow rate
1.5 to 2.3 LPM
Memory
21,500 data points
Locations
Up to 999 storage locations
Output RS-232
Operating temperature
32 to 120
°
F (0
°
-50
°
C)
Humidity range
95% non-condensing
Battery Rechargeable
NiCad
Battery life
8 hours
Charging time
8 hours
Size
9 x 3.5 x 2.5 in (22.9 x 8.9 x 6.4 cm)
Weight
3 lbs. (1.4 kg)
1 - 7