Pump setup, Keypad basics, Turning the pump on/off – SKC 100-3000 Leland Legacy User Manual
Page 9: Entering and navigating setup, Setup options
![background image](/manuals/349567/9/background.png)
6
Keypad Basics
Scrolls through run time data
and Setup options
Increases values such as fl ow
rate
Decreases values such as fl ow
rate
[]
When pressed simultaneously,
displayed item is selected or
entered.
Security code that must be
pressed in sequence to enter
Setup
Turning the Pump On/Off
• Press any butt on to turn on the power.
• Press [] to run the pump or to place a running pump in Hold.
• Manual Off : from Hold, press and hold .
• Auto Off turns off the pump aft er 5 minutes in Hold.
Entering and Navigating Setup
Entering:
Press [], then press the security code
in sequence. Setup should appear
briefl y on the LCD.
Navigating: Press to scroll through parameters. Once
the LCD shows End, parameters will repeat
until the user exits Setup.
Exiting:
Press until End appears on the LCD. Press
[]. The pump is now in Hold.
Setup Options
Aft er entering Setup, go to:
1. Flow
Set: Press or to increase or decrease pump
fl ow rate. Pump will start running. Press to move to
next parameter.
2. ADJ: Used during calibration with primary standard
calibrator (not for use with CalChek feature). Press
or to increase or decrease fl ow adjustment until
desired fl ow is indicated on calibrator. Press until End
appears. Press [] to save new fl ow and adjustment
sett ings and exit Setup.
Down
arrow
button
Button
Up
arrow
button
Pump Setup