Charging the Battery
Completely charge the batt ery pack before operating the pump. It may be
necessary to charge the batt ery a few times before maximum capacity is
achieved. Shown with single charger (Cat. No. 223-241). A fi ve-station charger
is available; see Optional Accessories on page 21. Follow charger instructions.
For a complete charge, ensure the pump is not running during charging.
• Use of a non-approved charger may damage the battery and pump and
VOIDS ANY WARRANTY.
• Using a repaired or rebuilt battery pack VOIDS ANY WARRANTY.
• Do not charge or operate pump with or without the charger in hazardous
locations.
• Ensure proper orientation of charging cable before plugging it into the
charging jack. Improper orientation/contact will short circuit the battery
and VOID ANY WARRANTY.
• Short circuiting the battery pack will render it immediately inoperative.
• Failure to follow warnings and cautions voids any warranty.
1. Insert the plug on the charging unit into the
batt ery charging jack on top of the pump
(underneath the protective cover).
2. Insert the plug on the power supply into the
jack on the charging unit.
3. Slide the appropriate wall plug into the
power supply and plug the power supply
into a wall outlet. The batt ery will recharge
in approximately 15 hours.
Aft er charging the batt ery pack, it is good practice to run the pump for
approximately 5 minutes before calibrating. This ensures the batt ery is
in more steady-state conditions and improves the agreement in pre and
post-sampling calibrations.
For more information on SKC pump batt ery packs,
The batt ery pack may be kept on the SKC-approved charger for an
indefi nite time.
Leland Legacy charging train with
single charger
Battery charging jack
Power
supply
Charging
unit
Jack