beautypg.com

Performance profi le – SKC 100-3000 Leland Legacy User Manual

Page 5

background image

2

Flow

Flow

Range:

5 to 15 L/min

Flow Control System:

Closed loop with patented* internal fl ow sensor

Compensation

Range:

15 L/min at 5 inches water back pressure

10 L/min at 12 inches water back pressure

5 L/min at 20 inches water back pressure

Typical Back Pressure of Sampling Media (inches water)

Compare the information in this table to pump compensation range to determine appropriate applications.

Accuracy:

Flow Rate: ± 5% of set-point after calibration to desired fl ow

Timing: 1 min/month at 25 C

Atmospheric

Pressure:

± 0.3 in Hg

Flow

Fault:

If the pump is unable to compensate for > 15 seconds due

to excessive back pressure, a fl ow fault icon displays and

fl ashes, the pump enters Hold mode, and the pump retains

historical data. Auto-restart is attempted every 20 seconds

up to 10 times. Adjustable with DataTrac for Leland Legacy

Software. See pages 18 and 20.

Tubing:

Requires 3/8-inch ID tubing

Operation

Display:

LCD displays pump serial number, pump software revision

level, flow rate, volume, temperature, atmospheric

pressure, time of day, run time, and pump status, i.e., Hold

and run as well as Setup information.

Volume

Display:

Continually updated, based on corrected fl ow rate

multiplied by sampling time. When volume exceeds

99,999 liters, the pump will continue to run normally but

an O_FLO Error will appear on the LCD.

Time

Display:

Time of day in hours and minutes (12 or 24-hour clock)

with AM and PM indicators

Timer Display Range:

1 to 99,999 minutes (69 days). If the run time exceeds

69 days, the timer display rolls over.

Operating Temp. Range: 32 to 113 F (0 to 45 C)

Protect sample pump from weather when

in use outdoors.

Altitude:

Do not use pump beyond 7500 ft.

Operating

Humidity:

0 to 95% non-condensing

Typical Run Time:

• Sioutas Impactor (approx. 13 in water bp): 24 hrs at

9 L/min

• PEM with 37-mm, 2.0-μm PTFE fi lter 225-1709

:

24 hrs at 10 L/min

• Low-volume PUF Tube: 24 hrs at 5 L/min

• DPS Sampler (PM2.5 or PM10): 24 hrs at 10 L/min

• 8 L/min Respirable PPI: 24 hrs at 8 L/min

• For extended run times, the pump may be operated

while attached to the approved charger.

Performance Profi le

Flow Rate (L/min)

5.0

8.0

10.0

12.0

15.0

Filter/Pore Size (μm)
37-mm MCE/0.8

11

18

22

28

36

37-mm PVC/5.0

4

7

9

11

15